Non-Invasive Blood Glucose Measurement
Megha C. Pande1,
Prof. A. K. Joshi2
1
Lecturer, Department of Instrumentation Engineering, COEM Manjari Pune
2
Asst. Professor Department of Instrumentation Engineering, CCOEW Pune
I. INTRODUCTION
Diabetes mellitus is a medical condition in which the body does not adequately produce the quantity or
the quality of insulin needed to maintain the normal circulating blood glucose. Insulin is the hormone that
enables glucose (Sugar) to enter the body cells to be used for energy. Two types of diabetes are common. Type
I is known as Insulin dependent diabetes Mellitus (IDDM) & accounts for 5-10% of all cases. Type II or Non-
insulin Dependent Diabetes Mellitus (NIDDM) requires insulin doses to maintain life, In addition to healthy
eating & exercise. NIDDM occurs later in life perhaps after 40 years of age and may require insulin or to be
controlled with an oral medication weight loss, a nutritious diet and a regular exercise.
Blood glucose measurements allow the detection of diabetic condition and are widely used by hospitals like
Operation Theatre, ICU, ER and Labour Room. [1]
A non-invasive blood glucose measurement has many advantages including the prevention of pain and
potential transmission of infectious diseases, reduced need for trained personnel, and relatively short
measurement time, and the absence of bio-hazardous waste. Since the near infrared light is found to penetrate a
great depth into biological tissues, near infrared spectroscopy has been used in development of non-invasive
method for biomedical sensing and clinical diagnosis. The absorption of whole blood in the visible and infrared
range is dominated by the different glucose derivatives and blood plasma that consists mainly of water. It is
well known that pulsatile changes of blood volume in tissue can be observed by measuring the transmission or
reflection of light through the blood volume.
Several methods are used to measure total blood glucose. The most common methods utilize
spectrophotometric analysis of light absorbance based on Beer –Lambert law.[2]
1. Methods for Glucose Measurement
1.1 Invasive Methods
Laboratory methods used for blood glucose testing
A) Reduction method
B) Condensation method
C) Enzymatic methods
1.2 Non Invasive Blood Glucose Measurement Techniques
Noninvasive methods mostly depend on absorption or transmittance. Blood glucose gives different
absorption spectrum for different wavelengths.
ABSTRACT:
The method for noninvasively blood glucose monitoring system is discussed in this paper. Lot of research
work has been done in developing the device which is completely non-invasive to avoid the pros & cons
because of frequent pricking. In this paper we are trying to analyse the non-invasive blood glucose
measurement study in the near infrared region which is the most suitable region for blood glucose
measurement. For this purpose we use a technique which is similar to pulse oximetry based on near
infrared spectrometry .An infrared light of particular wavelength is passed through fingertip containing
an arterial pulse component Thus minimizing influences of basal components such as resting blood
volume,skin, muscle and bone.
Keywords- Glucose measurement, Non-invasive, Photoplethysmography (PPG), Pulse-oximetry,
Infrared, Optical, Diabetes
Non-Invasive Blood Glucose…
www.ijceronline.com Open Access Journal Page 27
1.2.1 Optimal wavelength region in the near infrared [5]
The MIR light propagates only into a few scores or micrometer and may be applied for extracted blood
sample. On the other hand, the NIR light has deeper penetration into biological medium up to a few milimeters.
The NIR has a potential to be applied for non-invasive or minimally invasive blood analysis even though
glucose absorption is not as high as in the MIR region Following are the most common noninvasive methods.
1.2.2 Near infrared spectroscopy:
Glucose produces one of the weakest NIR absorption signal per concentration unit of the body major
component.NIR spectroscopy glucose measurement enables investigation of tissues depths in the range of 1 to
100 millimetres with a general decrease in penetration depth as wavelength value is increased.NIR transmission
through an earlobe, web and finger cuticle or reflected from the skin.
1.2.3 FIR spectroscopy:
A second technology for non-invasive blood glucose monitoring through spectroscope measures
absorption of FIR contained in natural thermal monitoring through spectroscope measures absorption of FIR
contained in natural thermal emission or body heat. FIR spectroscopy is the only type of radiation technology
that does not require an external energy source.
1.2.4 Raman spectroscopy:
Raman spectroscopy measures scattered light that has been influenced by the oscillation and rotation
of the scatter. Various Raman techniques have been attempted in blood, water, serum and plasma solutions.
Analytical problems include instability in the laser wavelengths and intensity & errors due to other chemicals in
the tissue sample and long spectral acquisition times [3]
1.2.5 Photo acoustic spectroscopy:
It uses optical beam to rapidly heat the sample and generate an acoustic pressure wave that can be
measured by microphone. The techniques are also subject to chemical interference from biological molecules as
well as physical interference from temperature and pressure changes. [4]
II. BASIC PRINCIPLE
Basic block diagram of noninvasive blood glucose measurement system is described in following
figure.The non-invasive sensor system allows a continuous measurement of the blood glucose concentration,
which is based on a pulse photometric measurement method. Thereby an area of skin is trans-illuminated by
light which is emitted by LEDs in the range SW NIR. Suitable wavelengths are selected for the analyses of
relative blood glucose concentration.
a. Mathematical implantation
Blood glucose is responsible for providing the energy to body. For spectrophotometric experiments,
Beer-lambert’s law is utilized and developed the notation of absorbance to express light absorption as a
function of blood glucose concentration given by following equation (1).
(1)
Where OD-Optical density
I0-Light intensity of incident light
I-Light intensity of transmitted light
-Extinction coefficient
C-Concentration of blood glucose
L-Length of light path through solution
b. Sensor Design
The developed glucose sensor system consist of number of hardware modules, which consist of
appropriate light sources, constant intensity circuit, detector circuit. The sensor consist of emitter LED’s with
center wavelength of simple RED light and 940nm.and 1450nm.comparative study for all wavelength is
performed on a specially designed hardware ckt each wavelength shows different kind of behaviour when
readings are taken with known concentration of glucose exposed to diff wavelengths. The LEDS are designed
to be placed opposite to a photo diode that detects the light from the LEDS. Absorption on each wavelength
differs significantly for different glucose concentration
Non-Invasive Blood Glucose…
www.ijceronline.com Open Access Journal Page 28
Detector ckt is common for all wavelengths and it is shown below:
Relationship with glucose concentration when experimentation is carried out with different glucose
concentration made from distilled water and sugar.
This single receiver photo diode is installed in the lower shell of finger clip. The probe is placed to the patient’s
body usually on the finger. The transmitted light is sensed by photodiode. Output voltage of photodiode varies
linearly with intensity.
III. RESULT AND DISCUSSIONS
An optical sensor is developed for blood glucose using wavelength of simple IR, 940nm and 1450nm.
Output is obtained on Digital storage oscilloscope Correlation is established between actual blood glucose
concentration and measured output voltage. Curve is fitted in fig. 1 and correlation is obtained From results it is
observed that 940nm and simple IR is giving the linear relationship .but 1450nm wavelength can be used for
non-invasive measurement since it has linear relationship with glucose concentration provided other error
causing elements are reduced. Still there is a scope for further development. A sensor developed with very high
signal to noise ratio can be designed and more no of readings taken would be an added advantage for future
scope of research in this field.
REFERENCES
[1] A. Caduff, F. Dewarrat, M. Talary, G. Stalder, L. Heinemann, and Y. Feldman, "Non-invasiveglucose
monitoring in patients with diabetes: Anovel system based on impedance spectroscopy,"
Biosens.Bioelectron., Mar.2006.
[2] A. Pfutzner, A. Caduff, M. Larbig, T. Schrepfer, and T. Forst, "Impact of posture and fixation technique
on impedance spectroscopy used for continuous and noninvasive glucose monitoring," Diabetes
Technol. Ther., vol. 6, no. 4, pp. 435-441, Aug.2004.
[3] T. Forst, A. Caduff, M. Talary, M. Weder, M.Brandle, P. Kann, F. Flacke, C. Friedrich, and A.Pfutzner,
"Impact of environmental temperature onskin thickness and microvascular blood flow insubjects with
and without diabetes," DiabetesTechnol. Ther., vol. 8, no. 1, pp. 94-101, Feb.2006.
[4] Heise, H.M. Medical applications of infrared spectroscopy. Mikrochim.Acta, in press, 1996.Wilson,
B.C. and Jacques, S.L. 0ptii:al reflectance and transmittance of tissue: principles and applications. IEEE
[5] Valery V. Tuchin (Ed.), Handbook of Optical Sensing of Glucose in Biological Fluids and Tissues
(CRC Press 2009 Taylor & Francis Group) 100-103.

More Related Content

What's hot

Ah atk2
Ah atk2Ah atk2
What is a Positron Emission Tomography?
What is a Positron Emission Tomography?What is a Positron Emission Tomography?
What is a Positron Emission Tomography?
Yvonne Saura
 
Exploring the Layered Structure of Composite Films Using an FTIR Microscope
Exploring the Layered Structure of Composite Films Using an FTIR MicroscopeExploring the Layered Structure of Composite Films Using an FTIR Microscope
Exploring the Layered Structure of Composite Films Using an FTIR Microscope
Shimadzu Scientific Instruments
 
Autoradiography
AutoradiographyAutoradiography
Autoradiography
PrakrutiKhatsuriya
 
Nirs
NirsNirs
Nirs
lundu
 
Flash radiation therapy
Flash radiation therapyFlash radiation therapy
Flash radiation therapy
Ajeet Gandhi
 
Illicit Drugs and Homemade Explosives Identification using FTIR Microscopy
Illicit Drugs and Homemade Explosives Identification using FTIR MicroscopyIllicit Drugs and Homemade Explosives Identification using FTIR Microscopy
Illicit Drugs and Homemade Explosives Identification using FTIR Microscopy
Shimadzu Scientific Instruments
 
Radioisotopes in biological system
Radioisotopes in biological systemRadioisotopes in biological system
Radioisotopes in biological system
zahra anwar
 
THE USE OF MICRO- AND MACRO-AUTORADIOGRAPHY TO STUDY THE TISSUE DISTRIBUTION ...
THE USE OF MICRO- AND MACRO-AUTORADIOGRAPHY TO STUDY THE TISSUE DISTRIBUTION ...THE USE OF MICRO- AND MACRO-AUTORADIOGRAPHY TO STUDY THE TISSUE DISTRIBUTION ...
THE USE OF MICRO- AND MACRO-AUTORADIOGRAPHY TO STUDY THE TISSUE DISTRIBUTION ...
QPS Holdings, LLC
 
Diffuse reflectance nir of plaque intracoronary device
Diffuse reflectance nir of plaque intracoronary deviceDiffuse reflectance nir of plaque intracoronary device
Diffuse reflectance nir of plaque intracoronary device
Society for Heart Attack Prevention and Eradication
 
Autoradiography
AutoradiographyAutoradiography
Autoradiography
Thippeswamy M
 
Photodynamic therapy and mechanism
Photodynamic therapy and mechanismPhotodynamic therapy and mechanism
Photodynamic therapy and mechanism
karoline Enoch
 
A NonInvasive Tool for the Screening of Neonatal Jaundice
A NonInvasive Tool for the Screening of Neonatal JaundiceA NonInvasive Tool for the Screening of Neonatal Jaundice
A NonInvasive Tool for the Screening of Neonatal Jaundice
avantika
 
Spectroscopic fiber optic catheter tania - aha 2001
Spectroscopic fiber optic catheter  tania - aha 2001Spectroscopic fiber optic catheter  tania - aha 2001
Spectroscopic fiber optic catheter tania - aha 2001
Society for Heart Attack Prevention and Eradication
 
094 spectroscopic fiber optic catheter
094 spectroscopic fiber optic catheter094 spectroscopic fiber optic catheter
094 spectroscopic fiber optic catheter
Society for Heart Attack Prevention and Eradication
 
Bio-optical Sensing Dissertation
Bio-optical Sensing DissertationBio-optical Sensing Dissertation
Bio-optical Sensing Dissertation
Ricky Hennessy
 
Research Progress of Phase Sensitive Optical Time Domain Vibration Sensing Te...
Research Progress of Phase Sensitive Optical Time Domain Vibration Sensing Te...Research Progress of Phase Sensitive Optical Time Domain Vibration Sensing Te...
Research Progress of Phase Sensitive Optical Time Domain Vibration Sensing Te...
IJRESJOURNAL
 
Micro autoradiography pptx
Micro autoradiography pptxMicro autoradiography pptx
Micro autoradiography pptx
ravi kiran
 
Various biochemistry lab instruments
Various biochemistry lab instrumentsVarious biochemistry lab instruments
Various biochemistry lab instruments
Kusum Jain
 

What's hot (19)

Ah atk2
Ah atk2Ah atk2
Ah atk2
 
What is a Positron Emission Tomography?
What is a Positron Emission Tomography?What is a Positron Emission Tomography?
What is a Positron Emission Tomography?
 
Exploring the Layered Structure of Composite Films Using an FTIR Microscope
Exploring the Layered Structure of Composite Films Using an FTIR MicroscopeExploring the Layered Structure of Composite Films Using an FTIR Microscope
Exploring the Layered Structure of Composite Films Using an FTIR Microscope
 
Autoradiography
AutoradiographyAutoradiography
Autoradiography
 
Nirs
NirsNirs
Nirs
 
Flash radiation therapy
Flash radiation therapyFlash radiation therapy
Flash radiation therapy
 
Illicit Drugs and Homemade Explosives Identification using FTIR Microscopy
Illicit Drugs and Homemade Explosives Identification using FTIR MicroscopyIllicit Drugs and Homemade Explosives Identification using FTIR Microscopy
Illicit Drugs and Homemade Explosives Identification using FTIR Microscopy
 
Radioisotopes in biological system
Radioisotopes in biological systemRadioisotopes in biological system
Radioisotopes in biological system
 
THE USE OF MICRO- AND MACRO-AUTORADIOGRAPHY TO STUDY THE TISSUE DISTRIBUTION ...
THE USE OF MICRO- AND MACRO-AUTORADIOGRAPHY TO STUDY THE TISSUE DISTRIBUTION ...THE USE OF MICRO- AND MACRO-AUTORADIOGRAPHY TO STUDY THE TISSUE DISTRIBUTION ...
THE USE OF MICRO- AND MACRO-AUTORADIOGRAPHY TO STUDY THE TISSUE DISTRIBUTION ...
 
Diffuse reflectance nir of plaque intracoronary device
Diffuse reflectance nir of plaque intracoronary deviceDiffuse reflectance nir of plaque intracoronary device
Diffuse reflectance nir of plaque intracoronary device
 
Autoradiography
AutoradiographyAutoradiography
Autoradiography
 
Photodynamic therapy and mechanism
Photodynamic therapy and mechanismPhotodynamic therapy and mechanism
Photodynamic therapy and mechanism
 
A NonInvasive Tool for the Screening of Neonatal Jaundice
A NonInvasive Tool for the Screening of Neonatal JaundiceA NonInvasive Tool for the Screening of Neonatal Jaundice
A NonInvasive Tool for the Screening of Neonatal Jaundice
 
Spectroscopic fiber optic catheter tania - aha 2001
Spectroscopic fiber optic catheter  tania - aha 2001Spectroscopic fiber optic catheter  tania - aha 2001
Spectroscopic fiber optic catheter tania - aha 2001
 
094 spectroscopic fiber optic catheter
094 spectroscopic fiber optic catheter094 spectroscopic fiber optic catheter
094 spectroscopic fiber optic catheter
 
Bio-optical Sensing Dissertation
Bio-optical Sensing DissertationBio-optical Sensing Dissertation
Bio-optical Sensing Dissertation
 
Research Progress of Phase Sensitive Optical Time Domain Vibration Sensing Te...
Research Progress of Phase Sensitive Optical Time Domain Vibration Sensing Te...Research Progress of Phase Sensitive Optical Time Domain Vibration Sensing Te...
Research Progress of Phase Sensitive Optical Time Domain Vibration Sensing Te...
 
Micro autoradiography pptx
Micro autoradiography pptxMicro autoradiography pptx
Micro autoradiography pptx
 
Various biochemistry lab instruments
Various biochemistry lab instrumentsVarious biochemistry lab instruments
Various biochemistry lab instruments
 

Similar to Non-Invasive Blood Glucose Measurement

U34129131
U34129131U34129131
U34129131
IJERA Editor
 
An embedded system for non invasive blood glucose measurement in glucose tole...
An embedded system for non invasive blood glucose measurement in glucose tole...An embedded system for non invasive blood glucose measurement in glucose tole...
An embedded system for non invasive blood glucose measurement in glucose tole...
eSAT Journals
 
Non-Invasive Blood Glucometer using NIR Spectroscopy
Non-Invasive Blood Glucometer using NIR Spectroscopy                            Non-Invasive Blood Glucometer using NIR Spectroscopy
Non-Invasive Blood Glucometer using NIR Spectroscopy
IRJET Journal
 
Effects of noises on near infrared sensor for blood glucose level measurement
Effects of noises on near infrared sensor for blood glucose level measurementEffects of noises on near infrared sensor for blood glucose level measurement
Effects of noises on near infrared sensor for blood glucose level measurement
TELKOMNIKA JOURNAL
 
Noninvasive blood glucose monitoring system based on near-infrared method
Noninvasive blood glucose monitoring system based on near-infrared method  Noninvasive blood glucose monitoring system based on near-infrared method
Noninvasive blood glucose monitoring system based on near-infrared method
IJECEIAES
 
Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...
Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...
Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...
TELKOMNIKA JOURNAL
 
LASER BASED NON-INVASIVE GLUCOSE MONITORING SYSTEM PPT.pptx
LASER BASED NON-INVASIVE GLUCOSE MONITORING SYSTEM PPT.pptxLASER BASED NON-INVASIVE GLUCOSE MONITORING SYSTEM PPT.pptx
LASER BASED NON-INVASIVE GLUCOSE MONITORING SYSTEM PPT.pptx
LokeshNivash1
 
A Non-Invasive Blood Glucose Monitoring Device using Red Laser Light
A Non-Invasive Blood Glucose Monitoring Device using Red Laser LightA Non-Invasive Blood Glucose Monitoring Device using Red Laser Light
A Non-Invasive Blood Glucose Monitoring Device using Red Laser Light
IRJET Journal
 
Implementation of multi-channel sensors for Bio-medical applications
Implementation of multi-channel sensors for Bio-medical applicationsImplementation of multi-channel sensors for Bio-medical applications
Implementation of multi-channel sensors for Bio-medical applications
Sree Bhargava Kothapalli
 
40120140507005
4012014050700540120140507005
40120140507005
IAEME Publication
 
40120140507005
4012014050700540120140507005
40120140507005
IAEME Publication
 
Melectronic pill
Melectronic pillMelectronic pill
Melectronic pill
punitkumar Kulli
 
biosensor and its utilization in environment
biosensor and its utilization in environmentbiosensor and its utilization in environment
biosensor and its utilization in environment
home
 
Diagnosis of chronic disease through biosensors copy
Diagnosis of chronic disease through biosensors   copyDiagnosis of chronic disease through biosensors   copy
Diagnosis of chronic disease through biosensors copy
mohammad shahrukh
 
A gsm based intelligent wireless mobile patient monitoring system
A gsm based intelligent wireless mobile patient monitoring systemA gsm based intelligent wireless mobile patient monitoring system
A gsm based intelligent wireless mobile patient monitoring system
eSAT Journals
 
International Journal of Computational Engineering Research(IJCER)
International Journal of Computational Engineering Research(IJCER) International Journal of Computational Engineering Research(IJCER)
International Journal of Computational Engineering Research(IJCER)
ijceronline
 
IRJET- Development of Portable Device for Measurement of Blood Glucose, T...
IRJET-  	  Development of Portable Device for Measurement of Blood Glucose, T...IRJET-  	  Development of Portable Device for Measurement of Blood Glucose, T...
IRJET- Development of Portable Device for Measurement of Blood Glucose, T...
IRJET Journal
 
Co32559562
Co32559562Co32559562
Co32559562
IJERA Editor
 
Merged document 12
Merged document 12Merged document 12
Merged document 12
V!vEk@nAnD S
 
IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...
IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...
IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...
IRJET Journal
 

Similar to Non-Invasive Blood Glucose Measurement (20)

U34129131
U34129131U34129131
U34129131
 
An embedded system for non invasive blood glucose measurement in glucose tole...
An embedded system for non invasive blood glucose measurement in glucose tole...An embedded system for non invasive blood glucose measurement in glucose tole...
An embedded system for non invasive blood glucose measurement in glucose tole...
 
Non-Invasive Blood Glucometer using NIR Spectroscopy
Non-Invasive Blood Glucometer using NIR Spectroscopy                            Non-Invasive Blood Glucometer using NIR Spectroscopy
Non-Invasive Blood Glucometer using NIR Spectroscopy
 
Effects of noises on near infrared sensor for blood glucose level measurement
Effects of noises on near infrared sensor for blood glucose level measurementEffects of noises on near infrared sensor for blood glucose level measurement
Effects of noises on near infrared sensor for blood glucose level measurement
 
Noninvasive blood glucose monitoring system based on near-infrared method
Noninvasive blood glucose monitoring system based on near-infrared method  Noninvasive blood glucose monitoring system based on near-infrared method
Noninvasive blood glucose monitoring system based on near-infrared method
 
Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...
Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...
Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...
 
LASER BASED NON-INVASIVE GLUCOSE MONITORING SYSTEM PPT.pptx
LASER BASED NON-INVASIVE GLUCOSE MONITORING SYSTEM PPT.pptxLASER BASED NON-INVASIVE GLUCOSE MONITORING SYSTEM PPT.pptx
LASER BASED NON-INVASIVE GLUCOSE MONITORING SYSTEM PPT.pptx
 
A Non-Invasive Blood Glucose Monitoring Device using Red Laser Light
A Non-Invasive Blood Glucose Monitoring Device using Red Laser LightA Non-Invasive Blood Glucose Monitoring Device using Red Laser Light
A Non-Invasive Blood Glucose Monitoring Device using Red Laser Light
 
Implementation of multi-channel sensors for Bio-medical applications
Implementation of multi-channel sensors for Bio-medical applicationsImplementation of multi-channel sensors for Bio-medical applications
Implementation of multi-channel sensors for Bio-medical applications
 
40120140507005
4012014050700540120140507005
40120140507005
 
40120140507005
4012014050700540120140507005
40120140507005
 
Melectronic pill
Melectronic pillMelectronic pill
Melectronic pill
 
biosensor and its utilization in environment
biosensor and its utilization in environmentbiosensor and its utilization in environment
biosensor and its utilization in environment
 
Diagnosis of chronic disease through biosensors copy
Diagnosis of chronic disease through biosensors   copyDiagnosis of chronic disease through biosensors   copy
Diagnosis of chronic disease through biosensors copy
 
A gsm based intelligent wireless mobile patient monitoring system
A gsm based intelligent wireless mobile patient monitoring systemA gsm based intelligent wireless mobile patient monitoring system
A gsm based intelligent wireless mobile patient monitoring system
 
International Journal of Computational Engineering Research(IJCER)
International Journal of Computational Engineering Research(IJCER) International Journal of Computational Engineering Research(IJCER)
International Journal of Computational Engineering Research(IJCER)
 
IRJET- Development of Portable Device for Measurement of Blood Glucose, T...
IRJET-  	  Development of Portable Device for Measurement of Blood Glucose, T...IRJET-  	  Development of Portable Device for Measurement of Blood Glucose, T...
IRJET- Development of Portable Device for Measurement of Blood Glucose, T...
 
Co32559562
Co32559562Co32559562
Co32559562
 
Merged document 12
Merged document 12Merged document 12
Merged document 12
 
IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...
IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...
IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...
 

Recently uploaded

Poznań ACE event - 19.06.2024 Team 24 Wrapup slidedeck
Poznań ACE event - 19.06.2024 Team 24 Wrapup slidedeckPoznań ACE event - 19.06.2024 Team 24 Wrapup slidedeck
Poznań ACE event - 19.06.2024 Team 24 Wrapup slidedeck
FilipTomaszewski5
 
Mutation Testing for Task-Oriented Chatbots
Mutation Testing for Task-Oriented ChatbotsMutation Testing for Task-Oriented Chatbots
Mutation Testing for Task-Oriented Chatbots
Pablo Gómez Abajo
 
Call Girls Chandigarh🔥7023059433🔥Agency Profile Escorts in Chandigarh Availab...
Call Girls Chandigarh🔥7023059433🔥Agency Profile Escorts in Chandigarh Availab...Call Girls Chandigarh🔥7023059433🔥Agency Profile Escorts in Chandigarh Availab...
Call Girls Chandigarh🔥7023059433🔥Agency Profile Escorts in Chandigarh Availab...
manji sharman06
 
PRODUCT LISTING OPTIMIZATION PRESENTATION.pptx
PRODUCT LISTING OPTIMIZATION PRESENTATION.pptxPRODUCT LISTING OPTIMIZATION PRESENTATION.pptx
PRODUCT LISTING OPTIMIZATION PRESENTATION.pptx
christinelarrosa
 
"Frontline Battles with DDoS: Best practices and Lessons Learned", Igor Ivaniuk
"Frontline Battles with DDoS: Best practices and Lessons Learned",  Igor Ivaniuk"Frontline Battles with DDoS: Best practices and Lessons Learned",  Igor Ivaniuk
"Frontline Battles with DDoS: Best practices and Lessons Learned", Igor Ivaniuk
Fwdays
 
MySQL InnoDB Storage Engine: Deep Dive - Mydbops
MySQL InnoDB Storage Engine: Deep Dive - MydbopsMySQL InnoDB Storage Engine: Deep Dive - Mydbops
MySQL InnoDB Storage Engine: Deep Dive - Mydbops
Mydbops
 
[OReilly Superstream] Occupy the Space: A grassroots guide to engineering (an...
[OReilly Superstream] Occupy the Space: A grassroots guide to engineering (an...[OReilly Superstream] Occupy the Space: A grassroots guide to engineering (an...
[OReilly Superstream] Occupy the Space: A grassroots guide to engineering (an...
Jason Yip
 
Multivendor cloud production with VSF TR-11 - there and back again
Multivendor cloud production with VSF TR-11 - there and back againMultivendor cloud production with VSF TR-11 - there and back again
Multivendor cloud production with VSF TR-11 - there and back again
Kieran Kunhya
 
Christine's Product Research Presentation.pptx
Christine's Product Research Presentation.pptxChristine's Product Research Presentation.pptx
Christine's Product Research Presentation.pptx
christinelarrosa
 
Must Know Postgres Extension for DBA and Developer during Migration
Must Know Postgres Extension for DBA and Developer during MigrationMust Know Postgres Extension for DBA and Developer during Migration
Must Know Postgres Extension for DBA and Developer during Migration
Mydbops
 
Northern Engraving | Modern Metal Trim, Nameplates and Appliance Panels
Northern Engraving | Modern Metal Trim, Nameplates and Appliance PanelsNorthern Engraving | Modern Metal Trim, Nameplates and Appliance Panels
Northern Engraving | Modern Metal Trim, Nameplates and Appliance Panels
Northern Engraving
 
Connector Corner: Seamlessly power UiPath Apps, GenAI with prebuilt connectors
Connector Corner: Seamlessly power UiPath Apps, GenAI with prebuilt connectorsConnector Corner: Seamlessly power UiPath Apps, GenAI with prebuilt connectors
Connector Corner: Seamlessly power UiPath Apps, GenAI with prebuilt connectors
DianaGray10
 
"Scaling RAG Applications to serve millions of users", Kevin Goedecke
"Scaling RAG Applications to serve millions of users",  Kevin Goedecke"Scaling RAG Applications to serve millions of users",  Kevin Goedecke
"Scaling RAG Applications to serve millions of users", Kevin Goedecke
Fwdays
 
Lee Barnes - Path to Becoming an Effective Test Automation Engineer.pdf
Lee Barnes - Path to Becoming an Effective Test Automation Engineer.pdfLee Barnes - Path to Becoming an Effective Test Automation Engineer.pdf
Lee Barnes - Path to Becoming an Effective Test Automation Engineer.pdf
leebarnesutopia
 
Northern Engraving | Nameplate Manufacturing Process - 2024
Northern Engraving | Nameplate Manufacturing Process - 2024Northern Engraving | Nameplate Manufacturing Process - 2024
Northern Engraving | Nameplate Manufacturing Process - 2024
Northern Engraving
 
A Deep Dive into ScyllaDB's Architecture
A Deep Dive into ScyllaDB's ArchitectureA Deep Dive into ScyllaDB's Architecture
A Deep Dive into ScyllaDB's Architecture
ScyllaDB
 
"Choosing proper type of scaling", Olena Syrota
"Choosing proper type of scaling", Olena Syrota"Choosing proper type of scaling", Olena Syrota
"Choosing proper type of scaling", Olena Syrota
Fwdays
 
Cyber Recovery Wargame
Cyber Recovery WargameCyber Recovery Wargame
Cyber Recovery Wargame
Databarracks
 
"NATO Hackathon Winner: AI-Powered Drug Search", Taras Kloba
"NATO Hackathon Winner: AI-Powered Drug Search",  Taras Kloba"NATO Hackathon Winner: AI-Powered Drug Search",  Taras Kloba
"NATO Hackathon Winner: AI-Powered Drug Search", Taras Kloba
Fwdays
 
Introducing BoxLang : A new JVM language for productivity and modularity!
Introducing BoxLang : A new JVM language for productivity and modularity!Introducing BoxLang : A new JVM language for productivity and modularity!
Introducing BoxLang : A new JVM language for productivity and modularity!
Ortus Solutions, Corp
 

Recently uploaded (20)

Poznań ACE event - 19.06.2024 Team 24 Wrapup slidedeck
Poznań ACE event - 19.06.2024 Team 24 Wrapup slidedeckPoznań ACE event - 19.06.2024 Team 24 Wrapup slidedeck
Poznań ACE event - 19.06.2024 Team 24 Wrapup slidedeck
 
Mutation Testing for Task-Oriented Chatbots
Mutation Testing for Task-Oriented ChatbotsMutation Testing for Task-Oriented Chatbots
Mutation Testing for Task-Oriented Chatbots
 
Call Girls Chandigarh🔥7023059433🔥Agency Profile Escorts in Chandigarh Availab...
Call Girls Chandigarh🔥7023059433🔥Agency Profile Escorts in Chandigarh Availab...Call Girls Chandigarh🔥7023059433🔥Agency Profile Escorts in Chandigarh Availab...
Call Girls Chandigarh🔥7023059433🔥Agency Profile Escorts in Chandigarh Availab...
 
PRODUCT LISTING OPTIMIZATION PRESENTATION.pptx
PRODUCT LISTING OPTIMIZATION PRESENTATION.pptxPRODUCT LISTING OPTIMIZATION PRESENTATION.pptx
PRODUCT LISTING OPTIMIZATION PRESENTATION.pptx
 
"Frontline Battles with DDoS: Best practices and Lessons Learned", Igor Ivaniuk
"Frontline Battles with DDoS: Best practices and Lessons Learned",  Igor Ivaniuk"Frontline Battles with DDoS: Best practices and Lessons Learned",  Igor Ivaniuk
"Frontline Battles with DDoS: Best practices and Lessons Learned", Igor Ivaniuk
 
MySQL InnoDB Storage Engine: Deep Dive - Mydbops
MySQL InnoDB Storage Engine: Deep Dive - MydbopsMySQL InnoDB Storage Engine: Deep Dive - Mydbops
MySQL InnoDB Storage Engine: Deep Dive - Mydbops
 
[OReilly Superstream] Occupy the Space: A grassroots guide to engineering (an...
[OReilly Superstream] Occupy the Space: A grassroots guide to engineering (an...[OReilly Superstream] Occupy the Space: A grassroots guide to engineering (an...
[OReilly Superstream] Occupy the Space: A grassroots guide to engineering (an...
 
Multivendor cloud production with VSF TR-11 - there and back again
Multivendor cloud production with VSF TR-11 - there and back againMultivendor cloud production with VSF TR-11 - there and back again
Multivendor cloud production with VSF TR-11 - there and back again
 
Christine's Product Research Presentation.pptx
Christine's Product Research Presentation.pptxChristine's Product Research Presentation.pptx
Christine's Product Research Presentation.pptx
 
Must Know Postgres Extension for DBA and Developer during Migration
Must Know Postgres Extension for DBA and Developer during MigrationMust Know Postgres Extension for DBA and Developer during Migration
Must Know Postgres Extension for DBA and Developer during Migration
 
Northern Engraving | Modern Metal Trim, Nameplates and Appliance Panels
Northern Engraving | Modern Metal Trim, Nameplates and Appliance PanelsNorthern Engraving | Modern Metal Trim, Nameplates and Appliance Panels
Northern Engraving | Modern Metal Trim, Nameplates and Appliance Panels
 
Connector Corner: Seamlessly power UiPath Apps, GenAI with prebuilt connectors
Connector Corner: Seamlessly power UiPath Apps, GenAI with prebuilt connectorsConnector Corner: Seamlessly power UiPath Apps, GenAI with prebuilt connectors
Connector Corner: Seamlessly power UiPath Apps, GenAI with prebuilt connectors
 
"Scaling RAG Applications to serve millions of users", Kevin Goedecke
"Scaling RAG Applications to serve millions of users",  Kevin Goedecke"Scaling RAG Applications to serve millions of users",  Kevin Goedecke
"Scaling RAG Applications to serve millions of users", Kevin Goedecke
 
Lee Barnes - Path to Becoming an Effective Test Automation Engineer.pdf
Lee Barnes - Path to Becoming an Effective Test Automation Engineer.pdfLee Barnes - Path to Becoming an Effective Test Automation Engineer.pdf
Lee Barnes - Path to Becoming an Effective Test Automation Engineer.pdf
 
Northern Engraving | Nameplate Manufacturing Process - 2024
Northern Engraving | Nameplate Manufacturing Process - 2024Northern Engraving | Nameplate Manufacturing Process - 2024
Northern Engraving | Nameplate Manufacturing Process - 2024
 
A Deep Dive into ScyllaDB's Architecture
A Deep Dive into ScyllaDB's ArchitectureA Deep Dive into ScyllaDB's Architecture
A Deep Dive into ScyllaDB's Architecture
 
"Choosing proper type of scaling", Olena Syrota
"Choosing proper type of scaling", Olena Syrota"Choosing proper type of scaling", Olena Syrota
"Choosing proper type of scaling", Olena Syrota
 
Cyber Recovery Wargame
Cyber Recovery WargameCyber Recovery Wargame
Cyber Recovery Wargame
 
"NATO Hackathon Winner: AI-Powered Drug Search", Taras Kloba
"NATO Hackathon Winner: AI-Powered Drug Search",  Taras Kloba"NATO Hackathon Winner: AI-Powered Drug Search",  Taras Kloba
"NATO Hackathon Winner: AI-Powered Drug Search", Taras Kloba
 
Introducing BoxLang : A new JVM language for productivity and modularity!
Introducing BoxLang : A new JVM language for productivity and modularity!Introducing BoxLang : A new JVM language for productivity and modularity!
Introducing BoxLang : A new JVM language for productivity and modularity!
 

Non-Invasive Blood Glucose Measurement

  • 1. ISSN (e): 2250 – 3005 || Volume, 05 || Issue, 04 || April – 2015 || International Journal of Computational Engineering Research (IJCER) www.ijceronline.com Open Access Journal Page 26 Non-Invasive Blood Glucose Measurement Megha C. Pande1, Prof. A. K. Joshi2 1 Lecturer, Department of Instrumentation Engineering, COEM Manjari Pune 2 Asst. Professor Department of Instrumentation Engineering, CCOEW Pune I. INTRODUCTION Diabetes mellitus is a medical condition in which the body does not adequately produce the quantity or the quality of insulin needed to maintain the normal circulating blood glucose. Insulin is the hormone that enables glucose (Sugar) to enter the body cells to be used for energy. Two types of diabetes are common. Type I is known as Insulin dependent diabetes Mellitus (IDDM) & accounts for 5-10% of all cases. Type II or Non- insulin Dependent Diabetes Mellitus (NIDDM) requires insulin doses to maintain life, In addition to healthy eating & exercise. NIDDM occurs later in life perhaps after 40 years of age and may require insulin or to be controlled with an oral medication weight loss, a nutritious diet and a regular exercise. Blood glucose measurements allow the detection of diabetic condition and are widely used by hospitals like Operation Theatre, ICU, ER and Labour Room. [1] A non-invasive blood glucose measurement has many advantages including the prevention of pain and potential transmission of infectious diseases, reduced need for trained personnel, and relatively short measurement time, and the absence of bio-hazardous waste. Since the near infrared light is found to penetrate a great depth into biological tissues, near infrared spectroscopy has been used in development of non-invasive method for biomedical sensing and clinical diagnosis. The absorption of whole blood in the visible and infrared range is dominated by the different glucose derivatives and blood plasma that consists mainly of water. It is well known that pulsatile changes of blood volume in tissue can be observed by measuring the transmission or reflection of light through the blood volume. Several methods are used to measure total blood glucose. The most common methods utilize spectrophotometric analysis of light absorbance based on Beer –Lambert law.[2] 1. Methods for Glucose Measurement 1.1 Invasive Methods Laboratory methods used for blood glucose testing A) Reduction method B) Condensation method C) Enzymatic methods 1.2 Non Invasive Blood Glucose Measurement Techniques Noninvasive methods mostly depend on absorption or transmittance. Blood glucose gives different absorption spectrum for different wavelengths. ABSTRACT: The method for noninvasively blood glucose monitoring system is discussed in this paper. Lot of research work has been done in developing the device which is completely non-invasive to avoid the pros & cons because of frequent pricking. In this paper we are trying to analyse the non-invasive blood glucose measurement study in the near infrared region which is the most suitable region for blood glucose measurement. For this purpose we use a technique which is similar to pulse oximetry based on near infrared spectrometry .An infrared light of particular wavelength is passed through fingertip containing an arterial pulse component Thus minimizing influences of basal components such as resting blood volume,skin, muscle and bone. Keywords- Glucose measurement, Non-invasive, Photoplethysmography (PPG), Pulse-oximetry, Infrared, Optical, Diabetes
  • 2. Non-Invasive Blood Glucose… www.ijceronline.com Open Access Journal Page 27 1.2.1 Optimal wavelength region in the near infrared [5] The MIR light propagates only into a few scores or micrometer and may be applied for extracted blood sample. On the other hand, the NIR light has deeper penetration into biological medium up to a few milimeters. The NIR has a potential to be applied for non-invasive or minimally invasive blood analysis even though glucose absorption is not as high as in the MIR region Following are the most common noninvasive methods. 1.2.2 Near infrared spectroscopy: Glucose produces one of the weakest NIR absorption signal per concentration unit of the body major component.NIR spectroscopy glucose measurement enables investigation of tissues depths in the range of 1 to 100 millimetres with a general decrease in penetration depth as wavelength value is increased.NIR transmission through an earlobe, web and finger cuticle or reflected from the skin. 1.2.3 FIR spectroscopy: A second technology for non-invasive blood glucose monitoring through spectroscope measures absorption of FIR contained in natural thermal monitoring through spectroscope measures absorption of FIR contained in natural thermal emission or body heat. FIR spectroscopy is the only type of radiation technology that does not require an external energy source. 1.2.4 Raman spectroscopy: Raman spectroscopy measures scattered light that has been influenced by the oscillation and rotation of the scatter. Various Raman techniques have been attempted in blood, water, serum and plasma solutions. Analytical problems include instability in the laser wavelengths and intensity & errors due to other chemicals in the tissue sample and long spectral acquisition times [3] 1.2.5 Photo acoustic spectroscopy: It uses optical beam to rapidly heat the sample and generate an acoustic pressure wave that can be measured by microphone. The techniques are also subject to chemical interference from biological molecules as well as physical interference from temperature and pressure changes. [4] II. BASIC PRINCIPLE Basic block diagram of noninvasive blood glucose measurement system is described in following figure.The non-invasive sensor system allows a continuous measurement of the blood glucose concentration, which is based on a pulse photometric measurement method. Thereby an area of skin is trans-illuminated by light which is emitted by LEDs in the range SW NIR. Suitable wavelengths are selected for the analyses of relative blood glucose concentration. a. Mathematical implantation Blood glucose is responsible for providing the energy to body. For spectrophotometric experiments, Beer-lambert’s law is utilized and developed the notation of absorbance to express light absorption as a function of blood glucose concentration given by following equation (1). (1) Where OD-Optical density I0-Light intensity of incident light I-Light intensity of transmitted light -Extinction coefficient C-Concentration of blood glucose L-Length of light path through solution b. Sensor Design The developed glucose sensor system consist of number of hardware modules, which consist of appropriate light sources, constant intensity circuit, detector circuit. The sensor consist of emitter LED’s with center wavelength of simple RED light and 940nm.and 1450nm.comparative study for all wavelength is performed on a specially designed hardware ckt each wavelength shows different kind of behaviour when readings are taken with known concentration of glucose exposed to diff wavelengths. The LEDS are designed to be placed opposite to a photo diode that detects the light from the LEDS. Absorption on each wavelength differs significantly for different glucose concentration
  • 3. Non-Invasive Blood Glucose… www.ijceronline.com Open Access Journal Page 28 Detector ckt is common for all wavelengths and it is shown below: Relationship with glucose concentration when experimentation is carried out with different glucose concentration made from distilled water and sugar. This single receiver photo diode is installed in the lower shell of finger clip. The probe is placed to the patient’s body usually on the finger. The transmitted light is sensed by photodiode. Output voltage of photodiode varies linearly with intensity. III. RESULT AND DISCUSSIONS An optical sensor is developed for blood glucose using wavelength of simple IR, 940nm and 1450nm. Output is obtained on Digital storage oscilloscope Correlation is established between actual blood glucose concentration and measured output voltage. Curve is fitted in fig. 1 and correlation is obtained From results it is observed that 940nm and simple IR is giving the linear relationship .but 1450nm wavelength can be used for non-invasive measurement since it has linear relationship with glucose concentration provided other error causing elements are reduced. Still there is a scope for further development. A sensor developed with very high signal to noise ratio can be designed and more no of readings taken would be an added advantage for future scope of research in this field. REFERENCES [1] A. Caduff, F. Dewarrat, M. Talary, G. Stalder, L. Heinemann, and Y. Feldman, "Non-invasiveglucose monitoring in patients with diabetes: Anovel system based on impedance spectroscopy," Biosens.Bioelectron., Mar.2006. [2] A. Pfutzner, A. Caduff, M. Larbig, T. Schrepfer, and T. Forst, "Impact of posture and fixation technique on impedance spectroscopy used for continuous and noninvasive glucose monitoring," Diabetes Technol. Ther., vol. 6, no. 4, pp. 435-441, Aug.2004. [3] T. Forst, A. Caduff, M. Talary, M. Weder, M.Brandle, P. Kann, F. Flacke, C. Friedrich, and A.Pfutzner, "Impact of environmental temperature onskin thickness and microvascular blood flow insubjects with and without diabetes," DiabetesTechnol. Ther., vol. 8, no. 1, pp. 94-101, Feb.2006. [4] Heise, H.M. Medical applications of infrared spectroscopy. Mikrochim.Acta, in press, 1996.Wilson, B.C. and Jacques, S.L. 0ptii:al reflectance and transmittance of tissue: principles and applications. IEEE [5] Valery V. Tuchin (Ed.), Handbook of Optical Sensing of Glucose in Biological Fluids and Tissues (CRC Press 2009 Taylor & Francis Group) 100-103.