Electrical Safety in Hospitals
Safety in Clinical Environment
• Electrical hazards
– Electrical shocks (micro and macro) due to equipment failure,
failure of power delivery systems, ground failures, burns, fire,
etc.
• Mechanical hazards
– mobility aids, transfer devices, prosthetic devices, mechanical
assist devices, patient support devices
• Environmental hazards
– Solid wastes, noise, utilities (natural gas), building structures,
etc.
• Biological hazards
– Infection control, viral outbreak, isolation, decontamination,
sterilization, waste disposal issues
• Radiation hazards
– Use of radioactive materials, radiation devices (MRI, CT, PET),
exposure control
Electrical Safety
• Many sources of energy, potentially hazardous substances,
instruments and procedures
– Use of fire, compressed air, water, chemicals, drugs,
microorganisms, waste, sound, electricity, radiation, natural and
unnatural disaster, negligence, sources of radiation, etc.
• Medical procedures expose patients to increased risks of
hazards due to skin and membranes being penetrated /
altered
• 10,000 device related injuries in the US every year!
Typically due to
– Improper use
– Inadequate training
– Lack of experience
– Improper (lack of) use of manuals
– Device failure
Physiological Effects of Electricity
• For electricity to have an effect on the human body:
– An electrical potential difference must be present
– The individual must be part of the electrical circuit, that is, a
current must enter the body at one point and leave it at
another.
• However, what causes the physiological effect is NOT
voltage, but rather CURRENT.
– A high voltage (KV,(103V)) applied over a large impedance
(rough skin) may not cause much (any) damage
– A low voltage applied over very small impedances (heart tissue)
may cause grave consequences (ventricular fibrillation)
• The magnitude of the current is simply the applied voltage
divided by the total effective impedance the current faces;
skin : largest.
• Electricity can have one of three effects:
– Electrical stimulation of excitable tissue (muscles, nerve)
– Resistive heating of tissue
– Electrical burns / tissue damage for direct current and high
voltages
Physiological Effects of Electricity
The real physiological effect depends on the actual path of the current
Physiological effects of electricity. Threshold or estimated mean values are given for each effect in a
70 kg human for a 1 to 3 s exposure to 60 Hz current applied via copper wires grasped by the hands.
Dry skin impedance:93 kΩ / cm2
Electrode gel on skin: 10.8 kΩ / cm2
Penetrated skin: 200 Ω / cm2
PHYSIOLOGICAL E FFECTS OF
E LECTRICITY
• Threshold of perception: The minimal current that an individual can detect. For
AC (with wet hands) can be as small as 0.5 mA at 60 Hz. For DC, 2 ~10 mA
• Let-go current: The maximal current at which the subject can voluntarily withdraw. 6 ~ 100 mA, at
which involuntary muscle contractions, reflex withdrawals, secondary physical effects (falling, hitting
head) may also occur
• Respiratory Paralysis / Pain / Fatigue At as low as 20 mA, involuntary
contractions of respiratory muscles can cause asphyxiation / respiratory arrest, if the current is not
interrupted. Strong involuntary contraction of other muscles can cause pain and fatigue
• Ventricular fibrillation 75 ~ 400 mA can cause heart muscles to contract
uncontrollably, altering the normal propagation of the electrical activity of the heart. HR can raise up
to 300 bpm, rapid, disorganized and too high to pump any meaningful amount of blood  ventricular
fibrillation. Normal rhythm can only return using a defibrillator
• Sustained myocardial contraction / Burns and physical
injury At 1 ~6 A, the entire heart muscle contracts and heart stops beating. This will not cause
irreversible tissue damage, however, as normal rhythm will return once the current is removed. At or
after 10A, however, burns can occur, particularly at points of entry and exit.
Important Susceptibility Parameters
• Threshold and let-go current variability
Distributions of perception thresholds and let-go currents These data depend on surface area of
contact, moistened hand graspingAWG No. 8 copper wire, 70 kg human, 60Hz, 1~3 s. of exposure
Important Susceptibility Parameters
e,
e
z.
• Frequency
– Note that the minimal
let-go current happens
at the precise
frequency of
commercial power-lin
50-60Hz.
Let-–
go c
L
ue
rre
tn
-t
gv
o
ers
c
uu
s f
rrr
eq
eu
n
en
tcy
rises
P
er
centbileevlaoluwesi
n
1d
0i
c
a
Ht
ezv
a
ar
i
a
nb
i
dl
i
t
y
ao
bf ov
let-go current among individuals. Let-go
currents
s f
e
or
vw
eo
rm
ae
lnh
ar
u
en
ab
d
ou
rt
etd
woH
-
thirds the values for men.
Important Susceptibility Parameters
• Duration
– The longer the duration, the
smaller the current at which
ventricular fibrillation occurs
– Shock must occur long
enough to coincide with the
most vulnerable period
occurring during the T wave.
• Weight
– Fibrillation threshold
increases with body weight
(from 50mA for 6kg dogs to
Fibrillat1io3n0cumrreAntfv
oe
r
rs
u
2s4s
h
o
kc
gkd
du
or
a
gt
i
so
.n
.
Thresholds for ventricular fibrillation in animals
for 60 Hz AC current. Duration of current (0.2 to
5 s) and weight of animal body were varied.
Important Susceptibility Parameters
• Points of entry
– The magnitude of the current required to fibrillate the heart is far
greater if the current is applied directly to heart; externally applied
current loses much of its amplitude due to current distributions. Large,
externally applied currents cause macroshock.
– If catheters are used, the natural protection provided by the skin 15 kΩ
~ 2 MΩ) is bypassed, greatly reducing the amount of current req’d to
cause fibrillation. Even smallest currents (80 ~ 600 μA), causing
microshock, may result in fibrillation. Safety limit for microshocks is 10
μA.
– The precise point of entry, even externally is very important: If both
points of entry and exit are on the same extremity, the risk of fibrillation
is greatly reduced even at high currents (e.g. the current req’d for
fibrillation through Lead I (LA-RA) electrodes is higher than for Leads II
(LL-RA) and III (LL-LA).
Important Susceptibility Parameters
• Points of entry
Effect of entry points on current distribution (a) Macroshock, externally applied current spreads through-
out the body. (b) Microshock, all the current applied through an intracardiac catheter flows through the
heart.
Macroshock
• Most electrical devices have a metal cabinet, which constitutes a hazard, in
case of an insulation failure or shortened component between the hot
power lead and the chassis. There is then 115 ~ 230 V between the chassis
and any other grounded object.
• The first line of defense available to patients is their skin.
– The outer layer provides 15 kΩ to 1 MΩ depending on the part of the body,
moisture and sweat present, 1% of that of dry skin if skin is broken,
– Internal resistance of the body is 200Ω for each limb, and 100Ω for the trunk,
thus internal body resistance between any two limbs is about 500Ω (somewhat
higher for obese people due to high resistivity of the adipose tissue)!
– Any procedure that reduces or eliminates the skin resistance increases the risk
of electrical shock, including biopotential electrode gel, electronic
thermometers placed in ears, mouth, rectum, intravenous catheters, etc.
• A third wire, grounded to earth, can greatly reduce the effect of
macroshock, as the resistance of that path would be much smaller then
even that of internal body resistance!
Macroshock Hazards
•Direct faults between the hot conductor and the ground is not
common,
and technically speaking, ground connection is not necessary during
normal operation.
•In fact, a ground fault will not be detected during normal operation of
the device, only when someone touches it, the hazard becomes
known.
Therefore, ground wire in devices and receptacles must be periodically
tested.
Macroshock Hazards
Microshock Hazards
Small currents inevitably flow between adjacent insulated conductors at different
potentials  leakagecurrentswhich flow through stray capacitances, insulation,
dust and moisture
Leakage current flowing to the chassis flows safely to the ground, if a low-resistance
ground wire is available.
Microshock Hazards
•If ground wire is broken, the chassis
potential rises above the ground; a
patient who has a grounded connection
to the heart (e.g. through a catheter)
receives a microshock if s/he touches
the chassis.
•If there is a connection from the
chassis to the patient’
s heart, anda
connection to the ground anywhere in
the body, this also causes microshock.
•Note that the hazard for microshock
only exists if there is a direct
connection to the heart. Otherwise,
even the internal resistance of the body
is high enough top prevent the
microshocks.
Microshock via
Ground Potentials
Microshocks can also occur if
different devices are not at the
exact same ground potential.
In fact, the microshock can
occur evenw
hena device that
does not connected to the patient
has a ground fault!
Afairly common ground
wire resistance of 0.1Ω can
easily cause a a 500mV
potential difference if
initiated due to a, say
5Aof ground fault.
If the patient resistance is less
then 50kΩ, this would cause an
above safe current of 10µA
Microshock Via
Ground Potentials
Safety Codes & Standards
• Limits on leakage current are instituted and
regulated by the safety codes instituted in part
by the National Fire Protection Association
(NFPA), American National Standards Institute
(ANSI), Association for the Advancement of
Medical Instrumentation (AAMI), and
Emergency Care Research Institute (ECRI).
Basic Approaches to
Shock Protection
• There are two major ways to protect patients from
shocks:
– Completely isolate and insulate patient from all sources of
electric current
– Keep all conductive surfaces within reach of the patient at
the same voltage
• Neither can be fully achieved  some combination of
these two
– Grounding system
– Isolated power-distribution system
– Ground-fault circuit interrupters (GFCI)
Grounding Systems
Low resistance (0.15 Ω) ground that can carry
currents up to the circuit-breaker ratings protects
patients by keeping all conductive surfaces and
receptacle grounds at the same potential.
Protects patients from
• Macroshocks
• Microshocks
• Ground faults elsewhere (!)
The difference between the receptacle grounds
and other surface should be no more then 40 mV)
All the receptacle grounds and conductive surfaces
in the vicinity of the patient are connected to the
patient-equipment grounding point. Each patient-
equipment grounding point is connected to the
reference grounding point that makes a single
connection to the building ground.
Isolated Power Systems
• A good equipotential grounding system cannot
eliminate large current that may result from major
ground-faults (which are rather rare).
• Isolated power systems can protect against such major
(single) ground faults
– Provide considerable protection against macroshocks,
particularly around wet conditions
– However, they are expensive !
– Used only at locations where flammable anesthetics are
used. Additional minor protection against microshocks
does not justify the high cost of these systems to be used
everywhere in the clinical environment
Protection through
Equipment Design
• Strain-relief devices for cords, where cord enters the
equipment and between the cord and plug
• Reduction of leakage current through proper layout
and insulation to minimize the capacitance between all
hot conductors and the chassis
• Double insulation to prevent the contact of the patient
with the chassis or any other conducting surface (outer
case being insulating material, plastic knobs, etc.)
• Operation at low voltages; solid state devices operating
at <10V are far less likely to cause macroshocks
• Electrical isolation in circuit design
Electrical Safety Analyzers
Testing Electrical Appliances
• Leakage current in patient leads:
– Potentially most damaging leakage is the one with patient
leads, since they typically have low impedance patient
contacts
– Current should be restricted to 50μA for non-isolated leads
and to 10 μA for isolated leads (used with catheters /
electrodes that make connection to the heart)
– Leakage current between any pair of leads, or between a
single lead and other patient connections should also be
controlled
– Leakage in case of line voltage appearing on the patient
should also be restricted.

More Related Content

Similar to ES.pptx

ELECTRICAL SAFETY pptx
ELECTRICAL SAFETY pptxELECTRICAL SAFETY pptx
ELECTRICAL SAFETY pptxmohamedrifky10
 
ELECTRICAL SAFETY IN OPERATION THEATRE .pptx
ELECTRICAL SAFETY IN OPERATION THEATRE .pptxELECTRICAL SAFETY IN OPERATION THEATRE .pptx
ELECTRICAL SAFETY IN OPERATION THEATRE .pptxSujata Walode
 
Electrical safety in OR
Electrical safety in ORElectrical safety in OR
Electrical safety in ORYamini Dhengle
 
Electrical Hazards and Patient Safety in Biomedical Equipment
Electrical Hazards and Patient Safety in Biomedical EquipmentElectrical Hazards and Patient Safety in Biomedical Equipment
Electrical Hazards and Patient Safety in Biomedical EquipmentGhansyam Rathod
 
Electrical safety training
Electrical safety trainingElectrical safety training
Electrical safety trainingTarkeeb
 
Report on Electrical Safety
Report on Electrical SafetyReport on Electrical Safety
Report on Electrical Safetyson_nikhil
 
electrical safety of bio-medical devices
electrical safety of bio-medical deviceselectrical safety of bio-medical devices
electrical safety of bio-medical devicesLongBun1
 
Electrical Burn.pptx
Electrical Burn.pptxElectrical Burn.pptx
Electrical Burn.pptxKishore Ravi
 
Electrical injuries
Electrical injuriesElectrical injuries
Electrical injuriesSudhir Dev
 
Electrical_Safety.pdf
Electrical_Safety.pdfElectrical_Safety.pdf
Electrical_Safety.pdfEmadTaleb1
 
She training module electrical safety
She training module electrical safetyShe training module electrical safety
She training module electrical safetysunilkumarkushwaha12
 
Surgical diathermy
Surgical diathermy Surgical diathermy
Surgical diathermy mdkaushar1
 
A17 Notes On Electrical Safety
A17 Notes On Electrical SafetyA17 Notes On Electrical Safety
A17 Notes On Electrical SafetyLisa Graves
 
Bio instrumentation
Bio instrumentation Bio instrumentation
Bio instrumentation mmZaeem
 

Similar to ES.pptx (20)

ELECTRICAL SAFETY pptx
ELECTRICAL SAFETY pptxELECTRICAL SAFETY pptx
ELECTRICAL SAFETY pptx
 
ELECTRICAL SAFETY IN OPERATION THEATRE .pptx
ELECTRICAL SAFETY IN OPERATION THEATRE .pptxELECTRICAL SAFETY IN OPERATION THEATRE .pptx
ELECTRICAL SAFETY IN OPERATION THEATRE .pptx
 
Electric burn injury- diagnosis and management
Electric burn injury- diagnosis and managementElectric burn injury- diagnosis and management
Electric burn injury- diagnosis and management
 
Electrical safety in OR
Electrical safety in ORElectrical safety in OR
Electrical safety in OR
 
Electrical Hazards and Patient Safety in Biomedical Equipment
Electrical Hazards and Patient Safety in Biomedical EquipmentElectrical Hazards and Patient Safety in Biomedical Equipment
Electrical Hazards and Patient Safety in Biomedical Equipment
 
Electrical safety training
Electrical safety trainingElectrical safety training
Electrical safety training
 
Electric shock
Electric shockElectric shock
Electric shock
 
Electric Shock-Ahdab
Electric Shock-AhdabElectric Shock-Ahdab
Electric Shock-Ahdab
 
Report on Electrical Safety
Report on Electrical SafetyReport on Electrical Safety
Report on Electrical Safety
 
electrical safety of bio-medical devices
electrical safety of bio-medical deviceselectrical safety of bio-medical devices
electrical safety of bio-medical devices
 
Electrical Burn.pptx
Electrical Burn.pptxElectrical Burn.pptx
Electrical Burn.pptx
 
Electrical injuries
Electrical injuriesElectrical injuries
Electrical injuries
 
Electrosurgery
ElectrosurgeryElectrosurgery
Electrosurgery
 
Electrical_Safety.pdf
Electrical_Safety.pdfElectrical_Safety.pdf
Electrical_Safety.pdf
 
She training module electrical safety
She training module electrical safetyShe training module electrical safety
She training module electrical safety
 
Electric Injuries
Electric InjuriesElectric Injuries
Electric Injuries
 
Surgical diathermy
Surgical diathermy Surgical diathermy
Surgical diathermy
 
A17 Notes On Electrical Safety
A17 Notes On Electrical SafetyA17 Notes On Electrical Safety
A17 Notes On Electrical Safety
 
Bio instrumentation
Bio instrumentation Bio instrumentation
Bio instrumentation
 
Energy sources in surgery
Energy sources in surgeryEnergy sources in surgery
Energy sources in surgery
 

More from ssuser1ecccc

Intro class. U1 M1.pptx
Intro class. U1 M1.pptxIntro class. U1 M1.pptx
Intro class. U1 M1.pptxssuser1ecccc
 
TASK Sixth Sensor Technology.pptx
TASK Sixth Sensor Technology.pptxTASK Sixth Sensor Technology.pptx
TASK Sixth Sensor Technology.pptxssuser1ecccc
 
4.-personal-productive-equipment.pptx
4.-personal-productive-equipment.pptx4.-personal-productive-equipment.pptx
4.-personal-productive-equipment.pptxssuser1ecccc
 
5.-accident-causation-theories-accident-reporting.pptx
5.-accident-causation-theories-accident-reporting.pptx5.-accident-causation-theories-accident-reporting.pptx
5.-accident-causation-theories-accident-reporting.pptxssuser1ecccc
 
unit-3-osha-hcs-dot.pptx
unit-3-osha-hcs-dot.pptxunit-3-osha-hcs-dot.pptx
unit-3-osha-hcs-dot.pptxssuser1ecccc
 
unit-iv-facility-safety.pptx
unit-iv-facility-safety.pptxunit-iv-facility-safety.pptx
unit-iv-facility-safety.pptxssuser1ecccc
 
MWM Unit 3 (1).pptx
MWM Unit 3 (1).pptxMWM Unit 3 (1).pptx
MWM Unit 3 (1).pptxssuser1ecccc
 
3.-unit-1-hazard-analysis-correction.pptx
3.-unit-1-hazard-analysis-correction.pptx3.-unit-1-hazard-analysis-correction.pptx
3.-unit-1-hazard-analysis-correction.pptxssuser1ecccc
 
Biomedical Waste Management.pptx
Biomedical Waste Management.pptxBiomedical Waste Management.pptx
Biomedical Waste Management.pptxssuser1ecccc
 
unit-2-major-minor-sources-of-biomedical-waste.pptx
unit-2-major-minor-sources-of-biomedical-waste.pptxunit-2-major-minor-sources-of-biomedical-waste.pptx
unit-2-major-minor-sources-of-biomedical-waste.pptxssuser1ecccc
 
Heart lung machine.pdf
Heart lung machine.pdfHeart lung machine.pdf
Heart lung machine.pdfssuser1ecccc
 
2.-unit-1-hazard-control-management-responsibility (1).pptx
2.-unit-1-hazard-control-management-responsibility (1).pptx2.-unit-1-hazard-control-management-responsibility (1).pptx
2.-unit-1-hazard-control-management-responsibility (1).pptxssuser1ecccc
 
heart-lung-machine-pump.ppt
heart-lung-machine-pump.pptheart-lung-machine-pump.ppt
heart-lung-machine-pump.pptssuser1ecccc
 
projectmanagement.ppt
projectmanagement.pptprojectmanagement.ppt
projectmanagement.pptssuser1ecccc
 

More from ssuser1ecccc (19)

Intro class. U1 M1.pptx
Intro class. U1 M1.pptxIntro class. U1 M1.pptx
Intro class. U1 M1.pptx
 
NGP BME Final.ppt
NGP BME Final.pptNGP BME Final.ppt
NGP BME Final.ppt
 
TASK Sixth Sensor Technology.pptx
TASK Sixth Sensor Technology.pptxTASK Sixth Sensor Technology.pptx
TASK Sixth Sensor Technology.pptx
 
4.-personal-productive-equipment.pptx
4.-personal-productive-equipment.pptx4.-personal-productive-equipment.pptx
4.-personal-productive-equipment.pptx
 
5.-accident-causation-theories-accident-reporting.pptx
5.-accident-causation-theories-accident-reporting.pptx5.-accident-causation-theories-accident-reporting.pptx
5.-accident-causation-theories-accident-reporting.pptx
 
unit-3-osha-hcs-dot.pptx
unit-3-osha-hcs-dot.pptxunit-3-osha-hcs-dot.pptx
unit-3-osha-hcs-dot.pptx
 
unit-iv-facility-safety.pptx
unit-iv-facility-safety.pptxunit-iv-facility-safety.pptx
unit-iv-facility-safety.pptx
 
MWM Unit 3 (1).pptx
MWM Unit 3 (1).pptxMWM Unit 3 (1).pptx
MWM Unit 3 (1).pptx
 
day9.ppt
day9.pptday9.ppt
day9.ppt
 
100014228.ppt
100014228.ppt100014228.ppt
100014228.ppt
 
3.-unit-1-hazard-analysis-correction.pptx
3.-unit-1-hazard-analysis-correction.pptx3.-unit-1-hazard-analysis-correction.pptx
3.-unit-1-hazard-analysis-correction.pptx
 
Biomedical Waste Management.pptx
Biomedical Waste Management.pptxBiomedical Waste Management.pptx
Biomedical Waste Management.pptx
 
UNIT 1 MWM.pptx
UNIT 1 MWM.pptxUNIT 1 MWM.pptx
UNIT 1 MWM.pptx
 
unit-2-major-minor-sources-of-biomedical-waste.pptx
unit-2-major-minor-sources-of-biomedical-waste.pptxunit-2-major-minor-sources-of-biomedical-waste.pptx
unit-2-major-minor-sources-of-biomedical-waste.pptx
 
Oxygenartors.pdf
Oxygenartors.pdfOxygenartors.pdf
Oxygenartors.pdf
 
Heart lung machine.pdf
Heart lung machine.pdfHeart lung machine.pdf
Heart lung machine.pdf
 
2.-unit-1-hazard-control-management-responsibility (1).pptx
2.-unit-1-hazard-control-management-responsibility (1).pptx2.-unit-1-hazard-control-management-responsibility (1).pptx
2.-unit-1-hazard-control-management-responsibility (1).pptx
 
heart-lung-machine-pump.ppt
heart-lung-machine-pump.pptheart-lung-machine-pump.ppt
heart-lung-machine-pump.ppt
 
projectmanagement.ppt
projectmanagement.pptprojectmanagement.ppt
projectmanagement.ppt
 

Recently uploaded

Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur Escorts
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur Suman Call 7001035870 Meet With Nagpur Escorts
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur High Profile
 
(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...
(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...
(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...ranjana rawat
 
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur Escorts
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur EscortsCall Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur Escorts
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur High Profile
 
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escortsranjana rawat
 
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINEMANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINESIVASHANKAR N
 
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service NashikCall Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service NashikCall Girls in Nagpur High Profile
 
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...ranjana rawat
 
UNIT-II FMM-Flow Through Circular Conduits
UNIT-II FMM-Flow Through Circular ConduitsUNIT-II FMM-Flow Through Circular Conduits
UNIT-II FMM-Flow Through Circular Conduitsrknatarajan
 
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...ranjana rawat
 
SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )Tsuyoshi Horigome
 
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escortsranjana rawat
 
Booking open Available Pune Call Girls Koregaon Park 6297143586 Call Hot Ind...
Booking open Available Pune Call Girls Koregaon Park  6297143586 Call Hot Ind...Booking open Available Pune Call Girls Koregaon Park  6297143586 Call Hot Ind...
Booking open Available Pune Call Girls Koregaon Park 6297143586 Call Hot Ind...Call Girls in Nagpur High Profile
 
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLSMANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLSSIVASHANKAR N
 
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...Soham Mondal
 
result management system report for college project
result management system report for college projectresult management system report for college project
result management system report for college projectTonystark477637
 
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).pptssuser5c9d4b1
 
Introduction to IEEE STANDARDS and its different types.pptx
Introduction to IEEE STANDARDS and its different types.pptxIntroduction to IEEE STANDARDS and its different types.pptx
Introduction to IEEE STANDARDS and its different types.pptxupamatechverse
 
Coefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptxCoefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptxAsutosh Ranjan
 

Recently uploaded (20)

Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur Escorts
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur Suman Call 7001035870 Meet With Nagpur Escorts
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur Escorts
 
(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...
(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...
(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...
 
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur Escorts
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur EscortsCall Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur Escorts
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur Escorts
 
Roadmap to Membership of RICS - Pathways and Routes
Roadmap to Membership of RICS - Pathways and RoutesRoadmap to Membership of RICS - Pathways and Routes
Roadmap to Membership of RICS - Pathways and Routes
 
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
 
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINEMANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
 
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service NashikCall Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
 
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
 
UNIT-II FMM-Flow Through Circular Conduits
UNIT-II FMM-Flow Through Circular ConduitsUNIT-II FMM-Flow Through Circular Conduits
UNIT-II FMM-Flow Through Circular Conduits
 
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
 
SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )
 
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
 
Booking open Available Pune Call Girls Koregaon Park 6297143586 Call Hot Ind...
Booking open Available Pune Call Girls Koregaon Park  6297143586 Call Hot Ind...Booking open Available Pune Call Girls Koregaon Park  6297143586 Call Hot Ind...
Booking open Available Pune Call Girls Koregaon Park 6297143586 Call Hot Ind...
 
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLSMANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
 
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
 
result management system report for college project
result management system report for college projectresult management system report for college project
result management system report for college project
 
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
 
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
 
Introduction to IEEE STANDARDS and its different types.pptx
Introduction to IEEE STANDARDS and its different types.pptxIntroduction to IEEE STANDARDS and its different types.pptx
Introduction to IEEE STANDARDS and its different types.pptx
 
Coefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptxCoefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptx
 

ES.pptx

  • 2. Safety in Clinical Environment • Electrical hazards – Electrical shocks (micro and macro) due to equipment failure, failure of power delivery systems, ground failures, burns, fire, etc. • Mechanical hazards – mobility aids, transfer devices, prosthetic devices, mechanical assist devices, patient support devices • Environmental hazards – Solid wastes, noise, utilities (natural gas), building structures, etc. • Biological hazards – Infection control, viral outbreak, isolation, decontamination, sterilization, waste disposal issues • Radiation hazards – Use of radioactive materials, radiation devices (MRI, CT, PET), exposure control
  • 3. Electrical Safety • Many sources of energy, potentially hazardous substances, instruments and procedures – Use of fire, compressed air, water, chemicals, drugs, microorganisms, waste, sound, electricity, radiation, natural and unnatural disaster, negligence, sources of radiation, etc. • Medical procedures expose patients to increased risks of hazards due to skin and membranes being penetrated / altered • 10,000 device related injuries in the US every year! Typically due to – Improper use – Inadequate training – Lack of experience – Improper (lack of) use of manuals – Device failure
  • 4. Physiological Effects of Electricity • For electricity to have an effect on the human body: – An electrical potential difference must be present – The individual must be part of the electrical circuit, that is, a current must enter the body at one point and leave it at another. • However, what causes the physiological effect is NOT voltage, but rather CURRENT. – A high voltage (KV,(103V)) applied over a large impedance (rough skin) may not cause much (any) damage – A low voltage applied over very small impedances (heart tissue) may cause grave consequences (ventricular fibrillation) • The magnitude of the current is simply the applied voltage divided by the total effective impedance the current faces; skin : largest. • Electricity can have one of three effects: – Electrical stimulation of excitable tissue (muscles, nerve) – Resistive heating of tissue – Electrical burns / tissue damage for direct current and high voltages
  • 5. Physiological Effects of Electricity The real physiological effect depends on the actual path of the current Physiological effects of electricity. Threshold or estimated mean values are given for each effect in a 70 kg human for a 1 to 3 s exposure to 60 Hz current applied via copper wires grasped by the hands. Dry skin impedance:93 kΩ / cm2 Electrode gel on skin: 10.8 kΩ / cm2 Penetrated skin: 200 Ω / cm2
  • 6. PHYSIOLOGICAL E FFECTS OF E LECTRICITY • Threshold of perception: The minimal current that an individual can detect. For AC (with wet hands) can be as small as 0.5 mA at 60 Hz. For DC, 2 ~10 mA • Let-go current: The maximal current at which the subject can voluntarily withdraw. 6 ~ 100 mA, at which involuntary muscle contractions, reflex withdrawals, secondary physical effects (falling, hitting head) may also occur • Respiratory Paralysis / Pain / Fatigue At as low as 20 mA, involuntary contractions of respiratory muscles can cause asphyxiation / respiratory arrest, if the current is not interrupted. Strong involuntary contraction of other muscles can cause pain and fatigue • Ventricular fibrillation 75 ~ 400 mA can cause heart muscles to contract uncontrollably, altering the normal propagation of the electrical activity of the heart. HR can raise up to 300 bpm, rapid, disorganized and too high to pump any meaningful amount of blood  ventricular fibrillation. Normal rhythm can only return using a defibrillator • Sustained myocardial contraction / Burns and physical injury At 1 ~6 A, the entire heart muscle contracts and heart stops beating. This will not cause irreversible tissue damage, however, as normal rhythm will return once the current is removed. At or after 10A, however, burns can occur, particularly at points of entry and exit.
  • 7. Important Susceptibility Parameters • Threshold and let-go current variability Distributions of perception thresholds and let-go currents These data depend on surface area of contact, moistened hand graspingAWG No. 8 copper wire, 70 kg human, 60Hz, 1~3 s. of exposure
  • 8. Important Susceptibility Parameters e, e z. • Frequency – Note that the minimal let-go current happens at the precise frequency of commercial power-lin 50-60Hz. Let-– go c L ue rre tn -t gv o ers c uu s f rrr eq eu n en tcy rises P er centbileevlaoluwesi n 1d 0i c a Ht ezv a ar i a nb i dl i t y ao bf ov let-go current among individuals. Let-go currents s f e or vw eo rm ae lnh ar u en ab d ou rt etd woH - thirds the values for men.
  • 9. Important Susceptibility Parameters • Duration – The longer the duration, the smaller the current at which ventricular fibrillation occurs – Shock must occur long enough to coincide with the most vulnerable period occurring during the T wave. • Weight – Fibrillation threshold increases with body weight (from 50mA for 6kg dogs to Fibrillat1io3n0cumrreAntfv oe r rs u 2s4s h o kc gkd du or a gt i so .n . Thresholds for ventricular fibrillation in animals for 60 Hz AC current. Duration of current (0.2 to 5 s) and weight of animal body were varied.
  • 10. Important Susceptibility Parameters • Points of entry – The magnitude of the current required to fibrillate the heart is far greater if the current is applied directly to heart; externally applied current loses much of its amplitude due to current distributions. Large, externally applied currents cause macroshock. – If catheters are used, the natural protection provided by the skin 15 kΩ ~ 2 MΩ) is bypassed, greatly reducing the amount of current req’d to cause fibrillation. Even smallest currents (80 ~ 600 μA), causing microshock, may result in fibrillation. Safety limit for microshocks is 10 μA. – The precise point of entry, even externally is very important: If both points of entry and exit are on the same extremity, the risk of fibrillation is greatly reduced even at high currents (e.g. the current req’d for fibrillation through Lead I (LA-RA) electrodes is higher than for Leads II (LL-RA) and III (LL-LA).
  • 11. Important Susceptibility Parameters • Points of entry Effect of entry points on current distribution (a) Macroshock, externally applied current spreads through- out the body. (b) Microshock, all the current applied through an intracardiac catheter flows through the heart.
  • 12. Macroshock • Most electrical devices have a metal cabinet, which constitutes a hazard, in case of an insulation failure or shortened component between the hot power lead and the chassis. There is then 115 ~ 230 V between the chassis and any other grounded object. • The first line of defense available to patients is their skin. – The outer layer provides 15 kΩ to 1 MΩ depending on the part of the body, moisture and sweat present, 1% of that of dry skin if skin is broken, – Internal resistance of the body is 200Ω for each limb, and 100Ω for the trunk, thus internal body resistance between any two limbs is about 500Ω (somewhat higher for obese people due to high resistivity of the adipose tissue)! – Any procedure that reduces or eliminates the skin resistance increases the risk of electrical shock, including biopotential electrode gel, electronic thermometers placed in ears, mouth, rectum, intravenous catheters, etc. • A third wire, grounded to earth, can greatly reduce the effect of macroshock, as the resistance of that path would be much smaller then even that of internal body resistance!
  • 13. Macroshock Hazards •Direct faults between the hot conductor and the ground is not common, and technically speaking, ground connection is not necessary during normal operation. •In fact, a ground fault will not be detected during normal operation of the device, only when someone touches it, the hazard becomes known. Therefore, ground wire in devices and receptacles must be periodically tested.
  • 15. Microshock Hazards Small currents inevitably flow between adjacent insulated conductors at different potentials  leakagecurrentswhich flow through stray capacitances, insulation, dust and moisture Leakage current flowing to the chassis flows safely to the ground, if a low-resistance ground wire is available.
  • 16. Microshock Hazards •If ground wire is broken, the chassis potential rises above the ground; a patient who has a grounded connection to the heart (e.g. through a catheter) receives a microshock if s/he touches the chassis. •If there is a connection from the chassis to the patient’ s heart, anda connection to the ground anywhere in the body, this also causes microshock. •Note that the hazard for microshock only exists if there is a direct connection to the heart. Otherwise, even the internal resistance of the body is high enough top prevent the microshocks.
  • 17. Microshock via Ground Potentials Microshocks can also occur if different devices are not at the exact same ground potential. In fact, the microshock can occur evenw hena device that does not connected to the patient has a ground fault! Afairly common ground wire resistance of 0.1Ω can easily cause a a 500mV potential difference if initiated due to a, say 5Aof ground fault. If the patient resistance is less then 50kΩ, this would cause an above safe current of 10µA
  • 19. Safety Codes & Standards • Limits on leakage current are instituted and regulated by the safety codes instituted in part by the National Fire Protection Association (NFPA), American National Standards Institute (ANSI), Association for the Advancement of Medical Instrumentation (AAMI), and Emergency Care Research Institute (ECRI).
  • 20. Basic Approaches to Shock Protection • There are two major ways to protect patients from shocks: – Completely isolate and insulate patient from all sources of electric current – Keep all conductive surfaces within reach of the patient at the same voltage • Neither can be fully achieved  some combination of these two – Grounding system – Isolated power-distribution system – Ground-fault circuit interrupters (GFCI)
  • 21. Grounding Systems Low resistance (0.15 Ω) ground that can carry currents up to the circuit-breaker ratings protects patients by keeping all conductive surfaces and receptacle grounds at the same potential. Protects patients from • Macroshocks • Microshocks • Ground faults elsewhere (!) The difference between the receptacle grounds and other surface should be no more then 40 mV) All the receptacle grounds and conductive surfaces in the vicinity of the patient are connected to the patient-equipment grounding point. Each patient- equipment grounding point is connected to the reference grounding point that makes a single connection to the building ground.
  • 22. Isolated Power Systems • A good equipotential grounding system cannot eliminate large current that may result from major ground-faults (which are rather rare). • Isolated power systems can protect against such major (single) ground faults – Provide considerable protection against macroshocks, particularly around wet conditions – However, they are expensive ! – Used only at locations where flammable anesthetics are used. Additional minor protection against microshocks does not justify the high cost of these systems to be used everywhere in the clinical environment
  • 23. Protection through Equipment Design • Strain-relief devices for cords, where cord enters the equipment and between the cord and plug • Reduction of leakage current through proper layout and insulation to minimize the capacitance between all hot conductors and the chassis • Double insulation to prevent the contact of the patient with the chassis or any other conducting surface (outer case being insulating material, plastic knobs, etc.) • Operation at low voltages; solid state devices operating at <10V are far less likely to cause macroshocks • Electrical isolation in circuit design
  • 24. Electrical Safety Analyzers Testing Electrical Appliances • Leakage current in patient leads: – Potentially most damaging leakage is the one with patient leads, since they typically have low impedance patient contacts – Current should be restricted to 50μA for non-isolated leads and to 10 μA for isolated leads (used with catheters / electrodes that make connection to the heart) – Leakage current between any pair of leads, or between a single lead and other patient connections should also be controlled – Leakage in case of line voltage appearing on the patient should also be restricted.