LCMS – Type of Scan
Chandra Prakash Singh
 The Quadrupole acts a mass filter; It separates ions based on m/z ratio.
 The Quadrupole is made of 4 rods:
 “A” pole refer to the vertical rods;
 “B” pole refer to the horizontal rods.
 DC & RF Voltage are imposed onto poles.
 RF and DC voltage are applied to “A” pole and “B” pole 180° out of phase
with each other.
 This happened so fast that ions never have chance to contact any one pole.
Quadrupole Theory
Ions are
attracted to a
“B” pole.
RF voltage
reversed, all
ions attracted
to an “A” pole.
RF voltage
reversed, all
ions attracted
to the other “B”
pole.
RF voltage
reversed, all
ions attracted
to other “A”
pole.
Principle of a Quadrupole:
 Two opposite quadrupole rods have
a positive, the other two a negative
polarity.
 The polarities of the two rod pairs
are exchanged with higher
frequency (RF) thus making ions
circulating between these four rods.
 A potential difference pushes the
circulating ions through the
quadrupole (ion energy).
Oscillating RF Effect on Ion Flight Path:
 At an RF/DC potential setting, only of one particular mass / charge ration have a stable trajectory.
 Ions which were produced in the source will be further << processed>> in quadrupole field.
A Quadrupole Can
Transfer ions
Scan ions
(sort according their m/z values)
Select ions with specified m/z
values
Stable and Unstable Ions Trajectories
m/z Selection:
Scan mode and Selected Ion Monitoring (SIM)
For a single quadrupole MS system, it can
operate in two modes:
Scan Mode
In scan mode, the voltages to the quadrupoles
is configured such that the entire mass range
specified is scanned sequentially with
appropriate dwell time for each m/z value.
Selected Ion Monitoring (SIM)
For the SIM mode Figure, only the
selected m/z (red ion) is monitored, passed
through the quadrupole and detected. This
SIM mode offers higher sensitivity and avoids
the effects of unwanted analytes and
impurities.
Scan Techniques
MS:
• In MS scans, also referred to as single MS scans, ions are separated according to their mass-to-charge ratio (m/z).
• A single MS scan might be used to determine the molecular weight of a compound.
• Single MS scans can also be referred to as survey scans.
• MS scans do not provide any information about the chemical make-up of the ions other than the mass.
MS/MS:
• MS/MS scans are used to determine a molecular species.
• For MS/MS scans in triple quadrupole systems, precursor ion fragmentation occurs in the collision cell.
• For MS/MS scans in QTRAP systems, precursor ion fragmentation can occur in the collision cell or the linear ion trap.
• If enough energy is used, then the precursor ion fragments to produce characteristic product ions.
MS/MS/MS:
• The linear ion trap (LIT) system MS/MS/MS scans go one step further than MS/MS scans.
• A fragment that is produced in the collision cell is fragmented further in the linear ion trap to give more structural
information about the molecular ion.
Quadrupole Scan Type
Q1/Q3 MS Information about a predefined masse range.
SIM (Q1/Q3 MI) Identification of molecular ions
MS/MS Structural Information, Library search
MRM Increased selectivity by “SIM” quantitation
Prec. / NL Increased selectivity, screening for defined analytes (e.g. metabolism studies)
MS and MS/MS Scan Type
MS Scans: MS/MS Scans:
 Q1/Q3 Full Range Scan
 Q1/Q3 Multiple Ion Scan
 Product Ion Scan
 Precursor Ion Scan
 Neutral Loss Scan
 Multiple Rection Monitoring (MRM)
MS Scan Type
MS Scan Range (Q1 or Q3): Multiple Ions (Q1 or Q3 MI):
 Full Range – Provide information for every
requested mass in the scan range.
 Or a center width scan – A narrow range scan to
obtain fast results for specific masses. This is used
for calibration and to confirm compound
characteristics.
 A scan using one quadrupole as the mass filter for specific
masses.
 Can be used for MS quantitation in addition to MS
compound parameter optimization.
Q1 MS (Q1 Scan):
Q1: Scan
The RF voltage combined with a DC
voltage stabilizes step by step one m/z
value at a given time. All other m/z
values are unstable and are lost.
 A full scan type using the first quadrupole (Q1).
 Monitor Present ions in a given mass Range.
 The ion intensity is returned for every mass in the scan range.
Q1 MS Scan:
Q2: RF Only
Q3: RF only
Q1 Multiple Ions (Q1 MI):
Q1: SIM
• A zero-width scan type using the Q1 quadrupole.
• The ion intensity is returned for the specified masses only.
• Use to optimized DP, EP, EP, CEP.
• The quantification is also possible. Q1 Multiple Ions
(MI – Single Ion Monitoring):
Q3: RF only Q2: RF Only
Q3 MS (Q3):
• A full scan type using the third quadrupole (Q3).
• Monitor present ions in a given mass range.
• The ion intensity is returned for every mass in the scan range.
Q3: Scan
The RF voltage combined with a
DC voltage stabilizes step by step
one m/z value at a given time. All
other m/z values are unstable and
are lost.
Q1: RF only Q2: RF only
Q3 Multiple Ions (Q3 MI): • A zero-width scan type using the Q3 quadrupole.
• The ion intensity is returned for the specified masses only.
Q1: RF only
Q2: RF only
Q3: SIM
Scan Type Q jet Qo Q1 Q2 Q3 Detector
Q1MS RF RF Scanning RF RF Detection
Q3MS RF RF RF RF Scanning Detection
Q1MI RF RF SIM RF RF Detection
Q3MI RF RF RF RF SIM/Fixed Detection
Ion Optics
Path
Orifice plate
QJet ion
guide
Q0 region
Q1
quadrupole
Q2 collision
cell
Q3
quadrupole
Detector
Quadrupole-Mode Scan Types
 Triple quadrupole instruments have high-sensitivity Multiple Reaction Monitoring (MRM) capabilities required for
quantitation experiments.
 In addition, they have highly specific scan types, such as precursor ion and neutral loss scans, which allow a more
advanced search to be performed on the components of the samples.
Fragmentation
 A charged precursor ion will decay into a charged and a neutral fragment:
 Secondary fragmentation may also take place.
 Fragmentation can take place at 2 locations:
 Interface
 Control via Declustering Potential (DP)
 Gas: CUR
 Collision Cell
 Control via Collision Energy (CE)
 Gas: CAD
MRM (Multiple Reaction Monitoring):
 Quantify multiple analytes by defining MRM-transitions (Q1 and Q3 pairs) in one experiment.
 The measure times (Dwell times) can be decreased up to 2 ms, with a minimum pause time of 3 ms.
 Twice as specific so that the quantification is more sensitive.
 Utilizing a MRM ion pair to quantify the analyte concentration of an unknown sample against a known concentration
calibration curve through regression analysis.
 An MS/MS scan in which a user-selected ion is passed
through the Q1 quadrupole and then fragmented in the
Q2 collision cell.
 The Q3 quadrupole then selects the fragment ion that
enters the detector.
 This scan mode is used primarily for quantitation.
 Substantial noise reduction by double mass filtering thus
increasing ratio S/N.
Q1: SIM
Q3:SIM
Q2: Fragment
MRM:
• An MS/MS full scan where the Q1 quadrupole is fixed to transmit a specific precursor ion and the Q3 quadrupole scans a
defined mass range.
• Used to identify all of the products of a particular precursor ion.
• Information about the structure of unknown compounds.
• Confirmation of known compounds.
• Library search
Product Ion (MS2):
This is when we know what the precursor or starting mass is.
Q1: Scan
Q2: Fragment
Q3: Scan
Product Ion (MS2) Scan:
 An MS/MS scan where the Q3 quadrupole is fixed at a specified mass-to-charge ratio to transmit a specific product ion
and the Q1 quadrupole scans a mass range.
 Used to confirm the presence of a precursor ion or, more commonly, to identify compounds sharing a common product
ion.
 Frequently used in the identification of drug metabolites (metabolites produce the same daughter ion).
Q1: Scan
Q2: Fragment
Q3: SIM
Filtering the mass of a
predefined charged fragment.
Search for
compound with the
same daughter ion.
Precursor Ion Scan (Prec):
Precursor Ion Scan (Prec):
Product Ion is 156.1 amu.
In this case, we would know what the product ion is ahead of time…
MS/MS – Precursor Ion Scan 1. Precursor ions scanned in Q1.
2. Fragmentation of precursor occurs in Q2
3. Product ion selected in Q3
Product ion is 156.1 amu
In this case, we would know what the product ion is ahead of time…
Neutral Loss Scan (NL)
 An MS/MS scan where both the Q1 quadrupole and the Q3 quadrupole scan a mass range, a fixed mass apart.
 A response is observed if the ion chosen by the Q1 quadrupole fragments by losing the neutral loss (the fixed mass)
specified.
 Used to confirm the presence of a precursor ion or, more commonly, to identify compounds sharing a common neutral loss.
Search for Compounds losing the same fragment.
Q1: Scan
Q2: Fragment
Q3: Scan Identifies the molecules that
generated a common neutral
fragment in the collision cell.
Parallel to Q1 with a predefined
mass difference which corresponds
to the mass of a neutral fragment.
Neutral Loss Scan
In this case you don’t know anything except that you have a neutral loss of a certain mass.
1. Precursor ions scanned in Q1.
2. Fragmentation of precursor occurs in Q2
3. Product ion selected in Q3
Neutral Loss Scan
Neutral is 95 amu.
In this case you don’t know anything except that
you have a neutral loss of a certain mass.
Scan Type Q jet Qo Q1 Q2 Q3 Detector
Q1MS RF RF Scanning RF RF Detection
Q3MS RF RF RF RF Scanning Detection
Q1MI RF RF SIM RF RF Detection
Q3MI RF RF RF RF SIM/Fixed Detection
Product Ion RF RF SIM Fragmentation Scanning Detection
MRM RF RF SIM Fragmentation SIM Detection
NL RF RF Scanning Fragmentation Scan Detection
Precursor Ion RF RF Scanning fragmentation SIM/Fixed Detection
Ion Optics
Path
Orifice plate
QJet ion
guide
Q0 region
Q1
quadrupole
Q2 collision
cell
Q3
quadrupole
Detector
 Ions are scanned in the Q1 quadrupole to the linear ion trap where they are collected.
 These ions are scanned out of the Q3 quadrupole to produce single MS type spectra.
Enhanced MS (EMS):
 The LIT-mode scans use the Q3 quadrupole as a linear ion trap.
 Ions are trapped and stored in the Q3 quadrupole before being scanned out, giving increased sensitivity.
 In addition, MS/MS/MS analysis can be performed in the linear ion trap, providing more information about the sample.
 LIT-mode scan types are typically used for qualitative measurements.
LIT-Mode Scan Types
 This scan type is similar to the EMS scan except that before ions are scanned out of the linear ion trap, there is a delay
period during which low-charge state ions (primarily singly-charged ions) are allowed to preferentially escape from the
linear ion trap.
 When the retained ions are scanned out, the multiply-charged ion population dominates the resulting spectrum.
Enhanced Multi-Charge (EMC):
 This scan type is used to obtain a high quality MS/MS spectrum on a specific ion.
 The fragmentation is done in the Q2 collision cell and thus provides the information-rich MS/MS spectrum typical of
collisionally activated dissociation (CAD) fragmentation.
 In this scan mode, the precursor ion to be fragmented is first selected in the Q1 quadrupole with a mass window that is 1
Da to 4 Da wide, filtering out all other ions.
 The precursor ion is fragmented by CAD gas in the Q2 collision cell. The fragment ions generated are captured in the
linear ion trap and then scanned out at one of three scan speeds, depending on the required fragment ion resolution.
 For IDA experiments, the Product Of field is set to 30 Da by default, and this value should not be changed.
Enhanced Product Ion (EPI):
 This scan type is similar to the EMS scan except that a small 30 Da mass around the precursor mass is scanned out of the
linear ion trap at the slowest scan rate to produce a narrow window of the best-resolved spectra.
Enhanced Resolution (ER):
 A precursor ion is selected by the Q1 quadrupole and fragmented with collisionally activated dissociation in the Q2
collision cell.
 The resulting product ions are all transmitted to the linear ion trap, where a single product ion is then isolated.
 The isolated ion is further fragmented in the linear ion trap, and the resulting product ions are scanned out of the trap at
one of three scan speeds.
 As with any in-trap Collision Induced Dissociation (CID) technique, there is a low mass cut-off for the second MS/MS
step due to the condition that the lowest mass fragment and precursor must be simultaneously stable in the trap.
 For QTRAP systems, this results in the loss of ions lower than 28 percent of the mass of the precursor ion during MS3
experiments.
 This phenomenon is often referred to as the one-third cut-off rule.
MS/MS/MS (MS3):
Scan Type Q jet Qo Q1 Q2 Q3 Detector
Q1MS RF RF Scanning RF RF Detection
EMS RF RF Scanning RF TRAP Detection
Q1MI RF RF SIM RF RF Detection
ER RF RF SIM RF TRAP Detection
Product Ion RF RF SIM Fragmentation SCAN Detection
EPI RF RF SIM Fragmentation TRAP Detection
EMC RF RF SCAN RF TRAP Detection
MRM RF RF SIM Fragmentation SIM Detection
MS/MS/MS RF RF SIM Fragmentation TRAP/ Fragmentation Detection
QTRAP
Retention Time: Time the peak elution from the LC. Typically expressed in minutes.
Scan Time (seconds): Time it takes for one pass across the mass range entered on one line of the mass range table.
Cycle (seconds): Time for one pass through the entries in the mass range table including the pause time between mass ranges.
Total Scan time (seconds): Total scan time for all entries in the mass range table(s) including pause time between mass range
and delay time.
Dwell Time (milliseconds): Represents the time the mass spectrometer monitors a single m/z or transition.
The Time (In MS)
LCMS Type of Scan.pdf

More Related Content

Similar to LCMS Type of Scan.pdf

Mass spectroscopy
Mass spectroscopyMass spectroscopy
Mass spectroscopylavanya0238
 
IR and NMR spectroscopy
IR and NMR spectroscopyIR and NMR spectroscopy
IR and NMR spectroscopySulov Saha
 
Types of spectrometer IMA
Types of spectrometer IMATypes of spectrometer IMA
Types of spectrometer IMAJAGESHWARSAHU2
 
liquid chromatography - mass spectroscopy (LC-MS)
liquid chromatography - mass spectroscopy (LC-MS)liquid chromatography - mass spectroscopy (LC-MS)
liquid chromatography - mass spectroscopy (LC-MS)akbar siddiq
 
MS/MS, Tandem Mass Spectrometry
MS/MS, Tandem Mass SpectrometryMS/MS, Tandem Mass Spectrometry
MS/MS, Tandem Mass SpectrometryVrushali Tambe
 
mass spectro pres.pptx
mass spectro pres.pptxmass spectro pres.pptx
mass spectro pres.pptxbreenaawan
 
Proton nmr by Bhushan Chavan
Proton nmr by Bhushan ChavanProton nmr by Bhushan Chavan
Proton nmr by Bhushan ChavanBhushan Chavan
 
quadrupole mass analyser Faiza hassan
quadrupole mass analyser Faiza hassanquadrupole mass analyser Faiza hassan
quadrupole mass analyser Faiza hassanfaiza bajwa
 
Mass spectrometry seminar
Mass spectrometry seminarMass spectrometry seminar
Mass spectrometry seminarPRIYANKA S
 
Nmr spectroscopy:- An overview and its principle
Nmr spectroscopy:- An overview and its principleNmr spectroscopy:- An overview and its principle
Nmr spectroscopy:- An overview and its principleSMGJAFAR
 
NMR Spectroscopy By Himaja Donthula
NMR Spectroscopy By Himaja DonthulaNMR Spectroscopy By Himaja Donthula
NMR Spectroscopy By Himaja Donthulahimaja donthula
 

Similar to LCMS Type of Scan.pdf (20)

Mass spectroscopy
Mass spectroscopyMass spectroscopy
Mass spectroscopy
 
Mass spectrometry
Mass spectrometryMass spectrometry
Mass spectrometry
 
Aadrsh kumar tiwari bbau
Aadrsh kumar tiwari bbauAadrsh kumar tiwari bbau
Aadrsh kumar tiwari bbau
 
Mass spectrometry new
Mass spectrometry newMass spectrometry new
Mass spectrometry new
 
IR and NMR spectroscopy
IR and NMR spectroscopyIR and NMR spectroscopy
IR and NMR spectroscopy
 
Types of spectrometer IMA
Types of spectrometer IMATypes of spectrometer IMA
Types of spectrometer IMA
 
Mass Analyser
Mass AnalyserMass Analyser
Mass Analyser
 
liquid chromatography - mass spectroscopy (LC-MS)
liquid chromatography - mass spectroscopy (LC-MS)liquid chromatography - mass spectroscopy (LC-MS)
liquid chromatography - mass spectroscopy (LC-MS)
 
MS/MS, Tandem Mass Spectrometry
MS/MS, Tandem Mass SpectrometryMS/MS, Tandem Mass Spectrometry
MS/MS, Tandem Mass Spectrometry
 
NMR Spectroscopy ppt.pptx
NMR Spectroscopy ppt.pptxNMR Spectroscopy ppt.pptx
NMR Spectroscopy ppt.pptx
 
mass spectro pres.pptx
mass spectro pres.pptxmass spectro pres.pptx
mass spectro pres.pptx
 
Proton nmr by Bhushan Chavan
Proton nmr by Bhushan ChavanProton nmr by Bhushan Chavan
Proton nmr by Bhushan Chavan
 
quadrupole mass analyser Faiza hassan
quadrupole mass analyser Faiza hassanquadrupole mass analyser Faiza hassan
quadrupole mass analyser Faiza hassan
 
Basic NMR
Basic NMRBasic NMR
Basic NMR
 
Infrared spectroscopy i
Infrared spectroscopy  iInfrared spectroscopy  i
Infrared spectroscopy i
 
Mass spectrometry seminar
Mass spectrometry seminarMass spectrometry seminar
Mass spectrometry seminar
 
Nmr spectroscopy:- An overview and its principle
Nmr spectroscopy:- An overview and its principleNmr spectroscopy:- An overview and its principle
Nmr spectroscopy:- An overview and its principle
 
Nuclear magnetic resonance
Nuclear magnetic resonanceNuclear magnetic resonance
Nuclear magnetic resonance
 
NMR Spectroscopy By Himaja Donthula
NMR Spectroscopy By Himaja DonthulaNMR Spectroscopy By Himaja Donthula
NMR Spectroscopy By Himaja Donthula
 
Nmr spectroscopy
Nmr spectroscopyNmr spectroscopy
Nmr spectroscopy
 

More from Chandra Prakash Singh

VALIDATION OF ANALYTICAL PROCEDURES Q2(R2) and Q14 - Terms and Definitions
VALIDATION OF ANALYTICAL PROCEDURES Q2(R2) and Q14 - Terms and DefinitionsVALIDATION OF ANALYTICAL PROCEDURES Q2(R2) and Q14 - Terms and Definitions
VALIDATION OF ANALYTICAL PROCEDURES Q2(R2) and Q14 - Terms and DefinitionsChandra Prakash Singh
 
VALIDATION OF ANALYTICAL PROCEDURES Q2(R2)
VALIDATION OF ANALYTICAL PROCEDURES Q2(R2)VALIDATION OF ANALYTICAL PROCEDURES Q2(R2)
VALIDATION OF ANALYTICAL PROCEDURES Q2(R2)Chandra Prakash Singh
 
NDSRIs - Nitrosamine Drug Substance-Related Impurities (NDSRIs)
NDSRIs - Nitrosamine Drug Substance-Related Impurities (NDSRIs)NDSRIs - Nitrosamine Drug Substance-Related Impurities (NDSRIs)
NDSRIs - Nitrosamine Drug Substance-Related Impurities (NDSRIs)Chandra Prakash Singh
 
Currently Identified Risk Factors for Presence of Nitrosamines.pptx
Currently Identified Risk Factors for Presence of Nitrosamines.pptxCurrently Identified Risk Factors for Presence of Nitrosamines.pptx
Currently Identified Risk Factors for Presence of Nitrosamines.pptxChandra Prakash Singh
 
X-Rays The basic understanding of XRD analysis
X-Rays The basic understanding of XRD analysisX-Rays The basic understanding of XRD analysis
X-Rays The basic understanding of XRD analysisChandra Prakash Singh
 
X-RAY POWDER DIFFRACTION (XRPD).pptx
X-RAY POWDER DIFFRACTION (XRPD).pptxX-RAY POWDER DIFFRACTION (XRPD).pptx
X-RAY POWDER DIFFRACTION (XRPD).pptxChandra Prakash Singh
 
LCMS - Ion Optics Path and Parameters.pptx
LCMS - Ion Optics Path and Parameters.pptxLCMS - Ion Optics Path and Parameters.pptx
LCMS - Ion Optics Path and Parameters.pptxChandra Prakash Singh
 
Analytical procedures life cycle management
Analytical procedures life cycle managementAnalytical procedures life cycle management
Analytical procedures life cycle managementChandra Prakash Singh
 
Acid base equilibria and reversed phase retention
Acid base equilibria and reversed phase retentionAcid base equilibria and reversed phase retention
Acid base equilibria and reversed phase retentionChandra Prakash Singh
 

More from Chandra Prakash Singh (20)

VALIDATION OF ANALYTICAL PROCEDURES Q2(R2) and Q14 - Terms and Definitions
VALIDATION OF ANALYTICAL PROCEDURES Q2(R2) and Q14 - Terms and DefinitionsVALIDATION OF ANALYTICAL PROCEDURES Q2(R2) and Q14 - Terms and Definitions
VALIDATION OF ANALYTICAL PROCEDURES Q2(R2) and Q14 - Terms and Definitions
 
VALIDATION OF ANALYTICAL PROCEDURES Q2(R2)
VALIDATION OF ANALYTICAL PROCEDURES Q2(R2)VALIDATION OF ANALYTICAL PROCEDURES Q2(R2)
VALIDATION OF ANALYTICAL PROCEDURES Q2(R2)
 
NDSRIs - Nitrosamine Drug Substance-Related Impurities (NDSRIs)
NDSRIs - Nitrosamine Drug Substance-Related Impurities (NDSRIs)NDSRIs - Nitrosamine Drug Substance-Related Impurities (NDSRIs)
NDSRIs - Nitrosamine Drug Substance-Related Impurities (NDSRIs)
 
Currently Identified Risk Factors for Presence of Nitrosamines.pptx
Currently Identified Risk Factors for Presence of Nitrosamines.pptxCurrently Identified Risk Factors for Presence of Nitrosamines.pptx
Currently Identified Risk Factors for Presence of Nitrosamines.pptx
 
X-Rays The basic understanding of XRD analysis
X-Rays The basic understanding of XRD analysisX-Rays The basic understanding of XRD analysis
X-Rays The basic understanding of XRD analysis
 
X-RAY POWDER DIFFRACTION (XRPD).pptx
X-RAY POWDER DIFFRACTION (XRPD).pptxX-RAY POWDER DIFFRACTION (XRPD).pptx
X-RAY POWDER DIFFRACTION (XRPD).pptx
 
Triplus500 HS working.pptx
Triplus500 HS working.pptxTriplus500 HS working.pptx
Triplus500 HS working.pptx
 
LCMS - Ion Optics Path and Parameters.pptx
LCMS - Ion Optics Path and Parameters.pptxLCMS - Ion Optics Path and Parameters.pptx
LCMS - Ion Optics Path and Parameters.pptx
 
Diverter Valve.pptx
Diverter Valve.pptxDiverter Valve.pptx
Diverter Valve.pptx
 
LCMS_interface.pptx
LCMS_interface.pptxLCMS_interface.pptx
LCMS_interface.pptx
 
Fitness for use part 1
Fitness for use part 1Fitness for use part 1
Fitness for use part 1
 
Analytical target profile 1
Analytical target profile 1Analytical target profile 1
Analytical target profile 1
 
Analytical control strategy 4
Analytical control strategy 4Analytical control strategy 4
Analytical control strategy 4
 
Analytical control strategy 3
Analytical control strategy 3Analytical control strategy 3
Analytical control strategy 3
 
Analytical control strategy 2
Analytical control strategy 2Analytical control strategy 2
Analytical control strategy 2
 
Analytical control strategy 1
Analytical control strategy 1Analytical control strategy 1
Analytical control strategy 1
 
Analytical procedures life cycle management
Analytical procedures life cycle managementAnalytical procedures life cycle management
Analytical procedures life cycle management
 
FID detector
FID detectorFID detector
FID detector
 
HPLC Column
HPLC ColumnHPLC Column
HPLC Column
 
Acid base equilibria and reversed phase retention
Acid base equilibria and reversed phase retentionAcid base equilibria and reversed phase retention
Acid base equilibria and reversed phase retention
 

Recently uploaded

mini mental status format.docx
mini    mental       status     format.docxmini    mental       status     format.docx
mini mental status format.docxPoojaSen20
 
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdfBASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdfSoniaTolstoy
 
Arihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdfArihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdfchloefrazer622
 
JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...
JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...
JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...anjaliyadav012327
 
Activity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfActivity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfciinovamais
 
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptxSOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptxiammrhaywood
 
Call Girls in Dwarka Mor Delhi Contact Us 9654467111
Call Girls in Dwarka Mor Delhi Contact Us 9654467111Call Girls in Dwarka Mor Delhi Contact Us 9654467111
Call Girls in Dwarka Mor Delhi Contact Us 9654467111Sapana Sha
 
Accessible design: Minimum effort, maximum impact
Accessible design: Minimum effort, maximum impactAccessible design: Minimum effort, maximum impact
Accessible design: Minimum effort, maximum impactdawncurless
 
Grant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy ConsultingGrant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy ConsultingTechSoup
 
Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)eniolaolutunde
 
Z Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot GraphZ Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot GraphThiyagu K
 
Organic Name Reactions for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions  for the students and aspirants of Chemistry12th.pptxOrganic Name Reactions  for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions for the students and aspirants of Chemistry12th.pptxVS Mahajan Coaching Centre
 
social pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajansocial pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajanpragatimahajan3
 
The byproduct of sericulture in different industries.pptx
The byproduct of sericulture in different industries.pptxThe byproduct of sericulture in different industries.pptx
The byproduct of sericulture in different industries.pptxShobhayan Kirtania
 
Paris 2024 Olympic Geographies - an activity
Paris 2024 Olympic Geographies - an activityParis 2024 Olympic Geographies - an activity
Paris 2024 Olympic Geographies - an activityGeoBlogs
 
Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...
Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...
Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...fonyou31
 
Introduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The BasicsIntroduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The BasicsTechSoup
 
Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104misteraugie
 
A Critique of the Proposed National Education Policy Reform
A Critique of the Proposed National Education Policy ReformA Critique of the Proposed National Education Policy Reform
A Critique of the Proposed National Education Policy ReformChameera Dedduwage
 
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...Sapna Thakur
 

Recently uploaded (20)

mini mental status format.docx
mini    mental       status     format.docxmini    mental       status     format.docx
mini mental status format.docx
 
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdfBASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdf
 
Arihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdfArihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdf
 
JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...
JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...
JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...
 
Activity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfActivity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdf
 
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptxSOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
 
Call Girls in Dwarka Mor Delhi Contact Us 9654467111
Call Girls in Dwarka Mor Delhi Contact Us 9654467111Call Girls in Dwarka Mor Delhi Contact Us 9654467111
Call Girls in Dwarka Mor Delhi Contact Us 9654467111
 
Accessible design: Minimum effort, maximum impact
Accessible design: Minimum effort, maximum impactAccessible design: Minimum effort, maximum impact
Accessible design: Minimum effort, maximum impact
 
Grant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy ConsultingGrant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy Consulting
 
Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)
 
Z Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot GraphZ Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot Graph
 
Organic Name Reactions for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions  for the students and aspirants of Chemistry12th.pptxOrganic Name Reactions  for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions for the students and aspirants of Chemistry12th.pptx
 
social pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajansocial pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajan
 
The byproduct of sericulture in different industries.pptx
The byproduct of sericulture in different industries.pptxThe byproduct of sericulture in different industries.pptx
The byproduct of sericulture in different industries.pptx
 
Paris 2024 Olympic Geographies - an activity
Paris 2024 Olympic Geographies - an activityParis 2024 Olympic Geographies - an activity
Paris 2024 Olympic Geographies - an activity
 
Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...
Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...
Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...
 
Introduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The BasicsIntroduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The Basics
 
Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104
 
A Critique of the Proposed National Education Policy Reform
A Critique of the Proposed National Education Policy ReformA Critique of the Proposed National Education Policy Reform
A Critique of the Proposed National Education Policy Reform
 
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...
 

LCMS Type of Scan.pdf

  • 1. LCMS – Type of Scan Chandra Prakash Singh
  • 2.  The Quadrupole acts a mass filter; It separates ions based on m/z ratio.  The Quadrupole is made of 4 rods:  “A” pole refer to the vertical rods;  “B” pole refer to the horizontal rods.  DC & RF Voltage are imposed onto poles.  RF and DC voltage are applied to “A” pole and “B” pole 180° out of phase with each other.  This happened so fast that ions never have chance to contact any one pole. Quadrupole Theory
  • 3. Ions are attracted to a “B” pole. RF voltage reversed, all ions attracted to an “A” pole. RF voltage reversed, all ions attracted to the other “B” pole. RF voltage reversed, all ions attracted to other “A” pole. Principle of a Quadrupole:  Two opposite quadrupole rods have a positive, the other two a negative polarity.  The polarities of the two rod pairs are exchanged with higher frequency (RF) thus making ions circulating between these four rods.  A potential difference pushes the circulating ions through the quadrupole (ion energy). Oscillating RF Effect on Ion Flight Path:
  • 4.  At an RF/DC potential setting, only of one particular mass / charge ration have a stable trajectory.  Ions which were produced in the source will be further << processed>> in quadrupole field. A Quadrupole Can Transfer ions Scan ions (sort according their m/z values) Select ions with specified m/z values Stable and Unstable Ions Trajectories m/z Selection:
  • 5. Scan mode and Selected Ion Monitoring (SIM) For a single quadrupole MS system, it can operate in two modes: Scan Mode In scan mode, the voltages to the quadrupoles is configured such that the entire mass range specified is scanned sequentially with appropriate dwell time for each m/z value. Selected Ion Monitoring (SIM) For the SIM mode Figure, only the selected m/z (red ion) is monitored, passed through the quadrupole and detected. This SIM mode offers higher sensitivity and avoids the effects of unwanted analytes and impurities.
  • 6. Scan Techniques MS: • In MS scans, also referred to as single MS scans, ions are separated according to their mass-to-charge ratio (m/z). • A single MS scan might be used to determine the molecular weight of a compound. • Single MS scans can also be referred to as survey scans. • MS scans do not provide any information about the chemical make-up of the ions other than the mass. MS/MS: • MS/MS scans are used to determine a molecular species. • For MS/MS scans in triple quadrupole systems, precursor ion fragmentation occurs in the collision cell. • For MS/MS scans in QTRAP systems, precursor ion fragmentation can occur in the collision cell or the linear ion trap. • If enough energy is used, then the precursor ion fragments to produce characteristic product ions. MS/MS/MS: • The linear ion trap (LIT) system MS/MS/MS scans go one step further than MS/MS scans. • A fragment that is produced in the collision cell is fragmented further in the linear ion trap to give more structural information about the molecular ion.
  • 7. Quadrupole Scan Type Q1/Q3 MS Information about a predefined masse range. SIM (Q1/Q3 MI) Identification of molecular ions MS/MS Structural Information, Library search MRM Increased selectivity by “SIM” quantitation Prec. / NL Increased selectivity, screening for defined analytes (e.g. metabolism studies) MS and MS/MS Scan Type MS Scans: MS/MS Scans:  Q1/Q3 Full Range Scan  Q1/Q3 Multiple Ion Scan  Product Ion Scan  Precursor Ion Scan  Neutral Loss Scan  Multiple Rection Monitoring (MRM)
  • 8. MS Scan Type MS Scan Range (Q1 or Q3): Multiple Ions (Q1 or Q3 MI):  Full Range – Provide information for every requested mass in the scan range.  Or a center width scan – A narrow range scan to obtain fast results for specific masses. This is used for calibration and to confirm compound characteristics.  A scan using one quadrupole as the mass filter for specific masses.  Can be used for MS quantitation in addition to MS compound parameter optimization.
  • 9. Q1 MS (Q1 Scan): Q1: Scan The RF voltage combined with a DC voltage stabilizes step by step one m/z value at a given time. All other m/z values are unstable and are lost.  A full scan type using the first quadrupole (Q1).  Monitor Present ions in a given mass Range.  The ion intensity is returned for every mass in the scan range. Q1 MS Scan: Q2: RF Only Q3: RF only
  • 10. Q1 Multiple Ions (Q1 MI): Q1: SIM • A zero-width scan type using the Q1 quadrupole. • The ion intensity is returned for the specified masses only. • Use to optimized DP, EP, EP, CEP. • The quantification is also possible. Q1 Multiple Ions (MI – Single Ion Monitoring): Q3: RF only Q2: RF Only
  • 11. Q3 MS (Q3): • A full scan type using the third quadrupole (Q3). • Monitor present ions in a given mass range. • The ion intensity is returned for every mass in the scan range. Q3: Scan The RF voltage combined with a DC voltage stabilizes step by step one m/z value at a given time. All other m/z values are unstable and are lost. Q1: RF only Q2: RF only Q3 Multiple Ions (Q3 MI): • A zero-width scan type using the Q3 quadrupole. • The ion intensity is returned for the specified masses only. Q1: RF only Q2: RF only Q3: SIM
  • 12. Scan Type Q jet Qo Q1 Q2 Q3 Detector Q1MS RF RF Scanning RF RF Detection Q3MS RF RF RF RF Scanning Detection Q1MI RF RF SIM RF RF Detection Q3MI RF RF RF RF SIM/Fixed Detection Ion Optics Path Orifice plate QJet ion guide Q0 region Q1 quadrupole Q2 collision cell Q3 quadrupole Detector
  • 13. Quadrupole-Mode Scan Types  Triple quadrupole instruments have high-sensitivity Multiple Reaction Monitoring (MRM) capabilities required for quantitation experiments.  In addition, they have highly specific scan types, such as precursor ion and neutral loss scans, which allow a more advanced search to be performed on the components of the samples. Fragmentation  A charged precursor ion will decay into a charged and a neutral fragment:  Secondary fragmentation may also take place.  Fragmentation can take place at 2 locations:  Interface  Control via Declustering Potential (DP)  Gas: CUR  Collision Cell  Control via Collision Energy (CE)  Gas: CAD
  • 14. MRM (Multiple Reaction Monitoring):  Quantify multiple analytes by defining MRM-transitions (Q1 and Q3 pairs) in one experiment.  The measure times (Dwell times) can be decreased up to 2 ms, with a minimum pause time of 3 ms.  Twice as specific so that the quantification is more sensitive.  Utilizing a MRM ion pair to quantify the analyte concentration of an unknown sample against a known concentration calibration curve through regression analysis.  An MS/MS scan in which a user-selected ion is passed through the Q1 quadrupole and then fragmented in the Q2 collision cell.  The Q3 quadrupole then selects the fragment ion that enters the detector.  This scan mode is used primarily for quantitation.  Substantial noise reduction by double mass filtering thus increasing ratio S/N.
  • 16. • An MS/MS full scan where the Q1 quadrupole is fixed to transmit a specific precursor ion and the Q3 quadrupole scans a defined mass range. • Used to identify all of the products of a particular precursor ion. • Information about the structure of unknown compounds. • Confirmation of known compounds. • Library search Product Ion (MS2): This is when we know what the precursor or starting mass is. Q1: Scan Q2: Fragment Q3: Scan
  • 18.  An MS/MS scan where the Q3 quadrupole is fixed at a specified mass-to-charge ratio to transmit a specific product ion and the Q1 quadrupole scans a mass range.  Used to confirm the presence of a precursor ion or, more commonly, to identify compounds sharing a common product ion.  Frequently used in the identification of drug metabolites (metabolites produce the same daughter ion). Q1: Scan Q2: Fragment Q3: SIM Filtering the mass of a predefined charged fragment. Search for compound with the same daughter ion. Precursor Ion Scan (Prec):
  • 19. Precursor Ion Scan (Prec): Product Ion is 156.1 amu. In this case, we would know what the product ion is ahead of time…
  • 20. MS/MS – Precursor Ion Scan 1. Precursor ions scanned in Q1. 2. Fragmentation of precursor occurs in Q2 3. Product ion selected in Q3 Product ion is 156.1 amu In this case, we would know what the product ion is ahead of time…
  • 21. Neutral Loss Scan (NL)  An MS/MS scan where both the Q1 quadrupole and the Q3 quadrupole scan a mass range, a fixed mass apart.  A response is observed if the ion chosen by the Q1 quadrupole fragments by losing the neutral loss (the fixed mass) specified.  Used to confirm the presence of a precursor ion or, more commonly, to identify compounds sharing a common neutral loss. Search for Compounds losing the same fragment. Q1: Scan Q2: Fragment Q3: Scan Identifies the molecules that generated a common neutral fragment in the collision cell. Parallel to Q1 with a predefined mass difference which corresponds to the mass of a neutral fragment.
  • 22. Neutral Loss Scan In this case you don’t know anything except that you have a neutral loss of a certain mass.
  • 23. 1. Precursor ions scanned in Q1. 2. Fragmentation of precursor occurs in Q2 3. Product ion selected in Q3 Neutral Loss Scan Neutral is 95 amu. In this case you don’t know anything except that you have a neutral loss of a certain mass.
  • 24. Scan Type Q jet Qo Q1 Q2 Q3 Detector Q1MS RF RF Scanning RF RF Detection Q3MS RF RF RF RF Scanning Detection Q1MI RF RF SIM RF RF Detection Q3MI RF RF RF RF SIM/Fixed Detection Product Ion RF RF SIM Fragmentation Scanning Detection MRM RF RF SIM Fragmentation SIM Detection NL RF RF Scanning Fragmentation Scan Detection Precursor Ion RF RF Scanning fragmentation SIM/Fixed Detection Ion Optics Path Orifice plate QJet ion guide Q0 region Q1 quadrupole Q2 collision cell Q3 quadrupole Detector
  • 25.  Ions are scanned in the Q1 quadrupole to the linear ion trap where they are collected.  These ions are scanned out of the Q3 quadrupole to produce single MS type spectra. Enhanced MS (EMS):  The LIT-mode scans use the Q3 quadrupole as a linear ion trap.  Ions are trapped and stored in the Q3 quadrupole before being scanned out, giving increased sensitivity.  In addition, MS/MS/MS analysis can be performed in the linear ion trap, providing more information about the sample.  LIT-mode scan types are typically used for qualitative measurements. LIT-Mode Scan Types  This scan type is similar to the EMS scan except that before ions are scanned out of the linear ion trap, there is a delay period during which low-charge state ions (primarily singly-charged ions) are allowed to preferentially escape from the linear ion trap.  When the retained ions are scanned out, the multiply-charged ion population dominates the resulting spectrum. Enhanced Multi-Charge (EMC):
  • 26.  This scan type is used to obtain a high quality MS/MS spectrum on a specific ion.  The fragmentation is done in the Q2 collision cell and thus provides the information-rich MS/MS spectrum typical of collisionally activated dissociation (CAD) fragmentation.  In this scan mode, the precursor ion to be fragmented is first selected in the Q1 quadrupole with a mass window that is 1 Da to 4 Da wide, filtering out all other ions.  The precursor ion is fragmented by CAD gas in the Q2 collision cell. The fragment ions generated are captured in the linear ion trap and then scanned out at one of three scan speeds, depending on the required fragment ion resolution.  For IDA experiments, the Product Of field is set to 30 Da by default, and this value should not be changed. Enhanced Product Ion (EPI):  This scan type is similar to the EMS scan except that a small 30 Da mass around the precursor mass is scanned out of the linear ion trap at the slowest scan rate to produce a narrow window of the best-resolved spectra. Enhanced Resolution (ER):
  • 27.  A precursor ion is selected by the Q1 quadrupole and fragmented with collisionally activated dissociation in the Q2 collision cell.  The resulting product ions are all transmitted to the linear ion trap, where a single product ion is then isolated.  The isolated ion is further fragmented in the linear ion trap, and the resulting product ions are scanned out of the trap at one of three scan speeds.  As with any in-trap Collision Induced Dissociation (CID) technique, there is a low mass cut-off for the second MS/MS step due to the condition that the lowest mass fragment and precursor must be simultaneously stable in the trap.  For QTRAP systems, this results in the loss of ions lower than 28 percent of the mass of the precursor ion during MS3 experiments.  This phenomenon is often referred to as the one-third cut-off rule. MS/MS/MS (MS3):
  • 28. Scan Type Q jet Qo Q1 Q2 Q3 Detector Q1MS RF RF Scanning RF RF Detection EMS RF RF Scanning RF TRAP Detection Q1MI RF RF SIM RF RF Detection ER RF RF SIM RF TRAP Detection Product Ion RF RF SIM Fragmentation SCAN Detection EPI RF RF SIM Fragmentation TRAP Detection EMC RF RF SCAN RF TRAP Detection MRM RF RF SIM Fragmentation SIM Detection MS/MS/MS RF RF SIM Fragmentation TRAP/ Fragmentation Detection QTRAP
  • 29. Retention Time: Time the peak elution from the LC. Typically expressed in minutes. Scan Time (seconds): Time it takes for one pass across the mass range entered on one line of the mass range table. Cycle (seconds): Time for one pass through the entries in the mass range table including the pause time between mass ranges. Total Scan time (seconds): Total scan time for all entries in the mass range table(s) including pause time between mass range and delay time. Dwell Time (milliseconds): Represents the time the mass spectrometer monitors a single m/z or transition. The Time (In MS)