Dose Issues and
CT
Radiation
Radiation is the emission and
transmission of energy or particles of
matter from atoms that travels through
some material / through a vacuum1.
2 Types of Radiation
1. Particle Radiation (α & β
radiation)
2. Electromagnetic Radiation
(X and Gamma radiation)
Electromagnetic Radiation
http://www.mpoweruk.com/radio.htm
• Radiation Exposure: It describes the amount of radiation
traveling through the air. Many radiation monitors measure
exposure. The units for exposure are the roentgen (R)2 .
• Absorbed Dose: which describes the amount of radiation
energy deposited in the patient’s body as a result of
exposure. The units for absorbed dose are the radiation
absorbed dose (rad) and gray (Gy)2.
• Effective Dose: are calculated by calculating the
absorbed doses to individual organs weighted for their
radiation sensitivity.
• The radiation risk to the patient can be estimated from the
effective dose.
• The common unit to measure effective dose is the mSv 2.
Biological Effects of Radiation
Stochastic Effects:
• Stochastic effects are those that occur by chance
and consist primarily of cancer and genetic effects3.
• Stochastic effects often show up years after
exposure3.
• As the dose to an individual increases, the probability
that cancer or a genetic effect will occur also
increases3.
• There is no threshold dose below which it is
relatively certain that an adverse effect cannot occur
for Stochastic Effects 1.
Non Stochastic (Deterministic ) Effects:
• These are characterized by a threshold dose above
which they do occur (clear relationship between the
exposure and the effect)1.
• In addition; the magnitude of the effect is directly
proportional to the size of the dose4.
• Non stochastic effects typically result when very large
dosages of radiation are received in a short amount of
time4.
• Examples of non stochastic effects include erythema
(skin reddening), skin and tissue burns, cataract
formation, sterility, radiation sickness and death4.
CT’s Contribution
X-Ray Procedure Radiation
Dose
CT:
49%
Rest:
51%
Total X-Ray Procedures
CT:, 17,%
Rest:,
83,%
Estimated dosages of X-Ray
Procedures
Background radiation per annum ≈ 3mSv
1 CT Head ≈ 20 Skull X-ray Examinations
1 CT Abdomen/Pelvis (2 Phase Study) ≈ 32 Abdominal X-rays
1 CT Abdomen/Pelvis (Urogram) ≈ 60 Abdominal X-rays
1 CT Chest ≈ 70 Chest X-ray Examinations
Radiograph Effective Dose
(mSv)
DEXA Scan 0.001
Bite Wing Dental 0.004
Skull (AP&Lat) 0.1
Chest (AP&Lat) 0.1
Abdomen (AP) 0.7
T-Spine (AP&Lat) 1
L-Spine (AP&Lat) 1.5
CT Procedure Effective Dose
(mSv)
Head 2
Neck 3
Chest 7
Abdomen 8
Pelvis 6
T-Spine 6
Chest for PE 15
Coronary Angio 16
Factors to consider
4 Justification : Exam must be medically
• indicated.
5 Optimization: Do the least number of scans at the
lowest exposure .
6 Are non ionizing alternative better suited (MR
and Ultrasound)5.
1 Patient Age.
2 Patient Gender.
3 Patient size and weight.
CT Dose Descriptors
1. CT Dose Index (CTDIvol)
●This is the primary dose concept in MDCT measured
in milli Gray (mGy).
● It is directly displayed on the MDCT Scanner.
●CTDIvol (weighted average of dose across a single
slice) divided by a pitch factor.
●It can be used to directly compare
radiation dose from different scan
parameter settings.
But it doses not provide an accurate
dose to the patient and should not be used to
determine individual risk5.
CT Dose Descriptors
2. Dose Length Product (DLP)
●DLP is an indicator of the total dose for an entire CT
examination.
● DLP is determined by multiplying the CTDIvol value by
the scan length, (mGy-cm).
●DLP will change between patients due to differences
in their height and the resulting scan length.
●It also includes overscan (extra
Scanning from MDCT Scans).
●DLP alone is still not sufficient to
accurately asses the radiation risk to a patient
undergoing a MDCT scan5.
CT Dose Descriptors
3. Effective Dose (ED)
●ED reflects the radiation risk to the individual patient,
measured in milli Sievert (mSv).
●ED takes into account not only the amount of
radiation, but also the sensitivity of the tissue/organ
being radiated.
● ED = k x DLP
.
k is the weighting factor accounting for tissue
sensitivity.
●Still mathematical models to compute organ doses
are based on a standard adult (70kg) .
Thus ED estimation can underestimate the risk to
children/ thin patients and overestimate the risk to
obese patients5.
Factors Affecting Radiation Dose
1.Tube Current (mA)
●Tube Current in CT implies the
amount of x-rays produced in the x-ray
tube.
●mA and mAs has a linear
relationship with respect to CT
radiation dose.
●When all other technical factors are kept the same, and mAs
are lowered , the radiation dosed is also lowered6.
Factors Affecting Radiation Dose
2. Peak Tube Voltage (kVp)
●This is the peak potential
difference between the anode
and cathode in the x-ray tube.
●It cause acceleration of the
electrons toward the anode to
produce x-rays.
● The most common range of kVps used in CT are 80 -140.
●The output of an x-ray tube is approximately proportional
to kVp2 . Therefore if kVp is doubled the radiation dose
increase by a factor of 4 5.
Factors Affecting Radiation Dose
3. Pitch
●Pitch is defined as the ratio of table feed per gantry
rotation to total x-ray beam with.
● Radiation doses is inversely
Proportional to the pitch.
● Therefore the greater the pitch
the lower the radiation dose to the patient and vise-versa.
●Protocols demanding high spatial resolution (e.g. Cardiac
ct scans) an pitch < 1 is used, translating in high radiation
doses to the patient5.
Factors Affecting Radiation Dose
4. Filters
●Consists of absorbent material inserted
between the x-ray tube and the patient .
● Filters absorb low energy x-rays and
“harden” the x-ray beam.
● Low energy x-rays contribute disproportionally to patient
radiation dose (especially to skin dose).
●Bow tie filters (often used in CT) deliver a uniform
intensity x-ray beam across the scanning object, resulting
in a more uniform distribution of the dose5.
Factors Affecting Radiation Dose
5. Patient Centering
●Patients normally have an elliptical cross-
Section( thinner in the periphery and
ticker in the centre.
● Bowtie filters are thicker in the periphery
and thinner in the centre, resulting in more x-rays reaching
the ticker central parts and less the thinner periphery.
● However for these filters to function properly the patient
must be centred at the gantry iso-center.
●If the patient is positioned off centred less x-rays pass
through the patient’s thicker central part ,resulting in more
noise in the centre of the images. While more x-rays reach
the peripheral portion of the body cross-section, increasing
unwanted radiation dose6.
Factors Affecting Radiation Dose
6. Repeats
●Multiple Scans results in an increase in
radiation dose to the patient .
● During a single visit to a CT suite, the
patient may be scanned multiple times.
● Patients can also undergo multiple ct scans within a
short period of time.
●How many of these scans are necessary and appropriate
Is the critical question that needs to be answered5.
Factors Affecting Radiation Dose
7. Dose efficiency (Geometric efficiency)
●In MDCT it is required that an equal number of x-ray
photons reach all active detectors.
●To ensure this a
certain amount of over-
beaming, beyond the
active detectors are needed (called the penumbra).
●This proportion of wasted radiation and subsequent dose
penalty is more profound with ct protocols using a thinner
detector configuration than those using a thicker one .
●The narrower the beam and the less the active detector
elements , the more radiation dose will increase5.
Factors Affecting Radiation Dose
8. Angular (X/Y) Dynamic Dose Modulation
●It modulates the x-ray tube current during the scan
according to the size of the scanned
object to provide.
●This is to provide optimal
penetration of the x-ray beam at a
specific tube angle for a specific
object.
●It provides the best dose saving for structure that are not
uniform in the x/y planes.
●Provide dose saving of approximately 10% for head
scans and 20-40% for chest scans6.
Factors Affecting Radiation Dose
9. Longitudinal (Z-axis) Dose Modulation
●It modulates the mAs along the longitudinal (Z) axis.
●Information for this modulation
are obtained from the surview
(scout) scan.
● It reduce dose in organs that have
lower attenuation(e.g: lungs) due
to the lower dose needed to obtain
diagnostic images.
●It also maintains an uniform image quality within the
scan5.
Factors Affecting Radiation Dose
10. Cardiac ECG Dose Modulation
● Retrospectively gated cardiac cta’s.
- Higher radiation dose to patient.
- More information that may be diagnostically relevant.
- Give functional information (e.g. Ventricular ejection fractions)
● Prospectively gated cardiac cta’s.
- Less radiation dose to patient (up to 83% less than retrospective studies).
- Loss of information that may be diagnostically relevant.
- No functional information (e.g. Ventricular ejection fractions)
Factors Affecting Radiation Dose
11. Paediatric Protocols
●The goal is to obtain a balance
between image quality and dose.
● These scan must be done with
the minimum mA that will still
provide a diagnostic relevant set of images.
● Low kVp settings in patients <45kg can result in 30-70%
dose reduction.
●Doubling the pitch from 1 to 2 will result in a 50%
decrease in the radiation dose.
●Paediatric examinations should also be performed with the
shortest possible scan time (0.5 ms or less) to minimize
patient movement and thus avoid repeating scan
procedures7.
Smart Card Protection for
Patients
●Up until a decade ago, radiation protection
was centred around protecting staff at
medical facilities.
●Patient protection was seen as not so
important, seen that their exposure to
radiation was thought to be only a few times in a life time.
●This notion has certainly change, and the radiation dose
patients received have become more prominent.
●What's needed is a way to track/record the radiation doses
an individual received from medical facilities over his
lifetime.
●To accomplish this the IAEA has launched a smart card
REFERENCES
1. Health Physics Society
. 2013. http://hps.org/publicinformation/ate/faqs/whatisradiation.html [2
June 2014].
2. United States Nuclear Regulatory Commission. 2013. http://www.nrc.gov/about-
nrc/radiation/health-effects/measuring-radiation.html [02 June 2014].
3. United States Environmental Protection Agency
. 2012.
http://www.epa.gov/rpdweb00/understand/health_effects.html [02 June 2014].
4. NDT Resource Center. 2013. https://www.nde-
ed.org/EducationResources/CommunityCollege/RadiationSafety/biological/nonstochastic/nonst
ochastic.htm [02 June 2014]
5. Mahesh, M. 2009. MDCT Physics: The Basics-Technology, Image Quality and Radiation Dose.
Philadelphia: Lippincott Williams & Wilkins.
6. Kalra, M.K., Saini, S. & Rubin, G.D. 2008. MDCT from Protocols to Practice. Milan, Italy:
Springer-Verslag Italia.
7. Siegal, M.J. 2007. Pediatric Body CT. 2nd ed. Philadelphia: Lippincott Williams & Wilkins.
8. International Atomic Energy Agency
.
http://www.iaea.org/Publications/Magazines/Bulletin/Bull502/50205813137.html [02 June 2014]
The End

More Related Content

Similar to CT Dose Issues.pptx on the factors to be considered on radiation protection

Radiation Safety In The Cath Lab
Radiation Safety In The Cath LabRadiation Safety In The Cath Lab
Radiation Safety In The Cath Labmagdy elmasry
 
Radiation risk from medical imaging
Radiation risk from medical imagingRadiation risk from medical imaging
Radiation risk from medical imagingAmal Al-Yasiri
 
Medical uses of ionising radiation
Medical uses of ionising radiationMedical uses of ionising radiation
Medical uses of ionising radiationAmin Amin
 
IMRT and 3D CRT in cervical Cancers
IMRT and 3D CRT in cervical CancersIMRT and 3D CRT in cervical Cancers
IMRT and 3D CRT in cervical CancersSantam Chakraborty
 
Radiation Therapy 1 & 2
Radiation Therapy 1 & 2Radiation Therapy 1 & 2
Radiation Therapy 1 & 2SCDA
 
Dose Evaluation and Reporting
Dose Evaluation and Reporting Dose Evaluation and Reporting
Dose Evaluation and Reporting Carestream
 
EBRT and brachytherapy radiology go3.ppt
EBRT and brachytherapy radiology go3.pptEBRT and brachytherapy radiology go3.ppt
EBRT and brachytherapy radiology go3.pptHahLa2
 
ExamensPresentationDanielBjörkman
ExamensPresentationDanielBjörkmanExamensPresentationDanielBjörkman
ExamensPresentationDanielBjörkmanDaniel Björkman
 
SPECT/CT: HOW Much Radiation Dose CT Constitute
SPECT/CT: HOW Much Radiation Dose CT ConstituteSPECT/CT: HOW Much Radiation Dose CT Constitute
SPECT/CT: HOW Much Radiation Dose CT ConstituteShahid Younas
 
Patient data acquisition and treatment verification.pptx
Patient data acquisition and treatment verification.pptxPatient data acquisition and treatment verification.pptx
Patient data acquisition and treatment verification.pptxJuliyet K Joy
 
Physics of Nuclear Medicine, SPECT and PET.ppt
Physics of Nuclear Medicine, SPECT and PET.pptPhysics of Nuclear Medicine, SPECT and PET.ppt
Physics of Nuclear Medicine, SPECT and PET.pptHassan Chattha
 
IMRT in Head & Neck Cancer
IMRT in Head & Neck CancerIMRT in Head & Neck Cancer
IMRT in Head & Neck CancerJyotirup Goswami
 
Radiotherapy ppt. Types of radiation and chronic
Radiotherapy ppt. Types of radiation and chronicRadiotherapy ppt. Types of radiation and chronic
Radiotherapy ppt. Types of radiation and chronicneestom1998
 
medical radiation safety awareness
medical radiation safety awarenessmedical radiation safety awareness
medical radiation safety awarenessMmedsc Hahm
 
Radiology_Equipment_Lec-1_Dr. Emad Taleb.pdf
Radiology_Equipment_Lec-1_Dr. Emad Taleb.pdfRadiology_Equipment_Lec-1_Dr. Emad Taleb.pdf
Radiology_Equipment_Lec-1_Dr. Emad Taleb.pdfEmadTaleb1
 
Pre-Clinical-Radiation-Protection_-(002).pptx
Pre-Clinical-Radiation-Protection_-(002).pptxPre-Clinical-Radiation-Protection_-(002).pptx
Pre-Clinical-Radiation-Protection_-(002).pptxChuemmanuelndze
 
Pre-Clinical-Radiation-Protection_-(002).pptx
Pre-Clinical-Radiation-Protection_-(002).pptxPre-Clinical-Radiation-Protection_-(002).pptx
Pre-Clinical-Radiation-Protection_-(002).pptxmedoghoneam
 
Radiotherapy In Early Breast Cancer
Radiotherapy In Early Breast CancerRadiotherapy In Early Breast Cancer
Radiotherapy In Early Breast CancerDr.T.Sujit :-)
 

Similar to CT Dose Issues.pptx on the factors to be considered on radiation protection (20)

Radiation Safety In The Cath Lab
Radiation Safety In The Cath LabRadiation Safety In The Cath Lab
Radiation Safety In The Cath Lab
 
Radiation risk from medical imaging
Radiation risk from medical imagingRadiation risk from medical imaging
Radiation risk from medical imaging
 
Medical uses of ionising radiation
Medical uses of ionising radiationMedical uses of ionising radiation
Medical uses of ionising radiation
 
Srs and sbrt 2 dr.kiran
Srs and sbrt 2 dr.kiranSrs and sbrt 2 dr.kiran
Srs and sbrt 2 dr.kiran
 
IMRT and 3D CRT in cervical Cancers
IMRT and 3D CRT in cervical CancersIMRT and 3D CRT in cervical Cancers
IMRT and 3D CRT in cervical Cancers
 
Radiograph guidelines for orthodontists by Almuzian
Radiograph guidelines for orthodontists by AlmuzianRadiograph guidelines for orthodontists by Almuzian
Radiograph guidelines for orthodontists by Almuzian
 
Radiation Therapy 1 & 2
Radiation Therapy 1 & 2Radiation Therapy 1 & 2
Radiation Therapy 1 & 2
 
Dose Evaluation and Reporting
Dose Evaluation and Reporting Dose Evaluation and Reporting
Dose Evaluation and Reporting
 
EBRT and brachytherapy radiology go3.ppt
EBRT and brachytherapy radiology go3.pptEBRT and brachytherapy radiology go3.ppt
EBRT and brachytherapy radiology go3.ppt
 
ExamensPresentationDanielBjörkman
ExamensPresentationDanielBjörkmanExamensPresentationDanielBjörkman
ExamensPresentationDanielBjörkman
 
SPECT/CT: HOW Much Radiation Dose CT Constitute
SPECT/CT: HOW Much Radiation Dose CT ConstituteSPECT/CT: HOW Much Radiation Dose CT Constitute
SPECT/CT: HOW Much Radiation Dose CT Constitute
 
Patient data acquisition and treatment verification.pptx
Patient data acquisition and treatment verification.pptxPatient data acquisition and treatment verification.pptx
Patient data acquisition and treatment verification.pptx
 
Physics of Nuclear Medicine, SPECT and PET.ppt
Physics of Nuclear Medicine, SPECT and PET.pptPhysics of Nuclear Medicine, SPECT and PET.ppt
Physics of Nuclear Medicine, SPECT and PET.ppt
 
IMRT in Head & Neck Cancer
IMRT in Head & Neck CancerIMRT in Head & Neck Cancer
IMRT in Head & Neck Cancer
 
Radiotherapy ppt. Types of radiation and chronic
Radiotherapy ppt. Types of radiation and chronicRadiotherapy ppt. Types of radiation and chronic
Radiotherapy ppt. Types of radiation and chronic
 
medical radiation safety awareness
medical radiation safety awarenessmedical radiation safety awareness
medical radiation safety awareness
 
Radiology_Equipment_Lec-1_Dr. Emad Taleb.pdf
Radiology_Equipment_Lec-1_Dr. Emad Taleb.pdfRadiology_Equipment_Lec-1_Dr. Emad Taleb.pdf
Radiology_Equipment_Lec-1_Dr. Emad Taleb.pdf
 
Pre-Clinical-Radiation-Protection_-(002).pptx
Pre-Clinical-Radiation-Protection_-(002).pptxPre-Clinical-Radiation-Protection_-(002).pptx
Pre-Clinical-Radiation-Protection_-(002).pptx
 
Pre-Clinical-Radiation-Protection_-(002).pptx
Pre-Clinical-Radiation-Protection_-(002).pptxPre-Clinical-Radiation-Protection_-(002).pptx
Pre-Clinical-Radiation-Protection_-(002).pptx
 
Radiotherapy In Early Breast Cancer
Radiotherapy In Early Breast CancerRadiotherapy In Early Breast Cancer
Radiotherapy In Early Breast Cancer
 

Recently uploaded

Types of Journalistic Writing Grade 8.pptx
Types of Journalistic Writing Grade 8.pptxTypes of Journalistic Writing Grade 8.pptx
Types of Journalistic Writing Grade 8.pptxEyham Joco
 
Interactive Powerpoint_How to Master effective communication
Interactive Powerpoint_How to Master effective communicationInteractive Powerpoint_How to Master effective communication
Interactive Powerpoint_How to Master effective communicationnomboosow
 
Framing an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdf
Framing an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdfFraming an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdf
Framing an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdfUjwalaBharambe
 
ECONOMIC CONTEXT - LONG FORM TV DRAMA - PPT
ECONOMIC CONTEXT - LONG FORM TV DRAMA - PPTECONOMIC CONTEXT - LONG FORM TV DRAMA - PPT
ECONOMIC CONTEXT - LONG FORM TV DRAMA - PPTiammrhaywood
 
CARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxCARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxGaneshChakor2
 
Roles & Responsibilities in Pharmacovigilance
Roles & Responsibilities in PharmacovigilanceRoles & Responsibilities in Pharmacovigilance
Roles & Responsibilities in PharmacovigilanceSamikshaHamane
 
Proudly South Africa powerpoint Thorisha.pptx
Proudly South Africa powerpoint Thorisha.pptxProudly South Africa powerpoint Thorisha.pptx
Proudly South Africa powerpoint Thorisha.pptxthorishapillay1
 
Employee wellbeing at the workplace.pptx
Employee wellbeing at the workplace.pptxEmployee wellbeing at the workplace.pptx
Employee wellbeing at the workplace.pptxNirmalaLoungPoorunde1
 
“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...
“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...
“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...Marc Dusseiller Dusjagr
 
Painted Grey Ware.pptx, PGW Culture of India
Painted Grey Ware.pptx, PGW Culture of IndiaPainted Grey Ware.pptx, PGW Culture of India
Painted Grey Ware.pptx, PGW Culture of IndiaVirag Sontakke
 
Biting mechanism of poisonous snakes.pdf
Biting mechanism of poisonous snakes.pdfBiting mechanism of poisonous snakes.pdf
Biting mechanism of poisonous snakes.pdfadityarao40181
 
18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf
18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf
18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdfssuser54595a
 
How to Configure Email Server in Odoo 17
How to Configure Email Server in Odoo 17How to Configure Email Server in Odoo 17
How to Configure Email Server in Odoo 17Celine George
 
Presiding Officer Training module 2024 lok sabha elections
Presiding Officer Training module 2024 lok sabha electionsPresiding Officer Training module 2024 lok sabha elections
Presiding Officer Training module 2024 lok sabha electionsanshu789521
 
Historical philosophical, theoretical, and legal foundations of special and i...
Historical philosophical, theoretical, and legal foundations of special and i...Historical philosophical, theoretical, and legal foundations of special and i...
Historical philosophical, theoretical, and legal foundations of special and i...jaredbarbolino94
 
Computed Fields and api Depends in the Odoo 17
Computed Fields and api Depends in the Odoo 17Computed Fields and api Depends in the Odoo 17
Computed Fields and api Depends in the Odoo 17Celine George
 

Recently uploaded (20)

Types of Journalistic Writing Grade 8.pptx
Types of Journalistic Writing Grade 8.pptxTypes of Journalistic Writing Grade 8.pptx
Types of Journalistic Writing Grade 8.pptx
 
Interactive Powerpoint_How to Master effective communication
Interactive Powerpoint_How to Master effective communicationInteractive Powerpoint_How to Master effective communication
Interactive Powerpoint_How to Master effective communication
 
ESSENTIAL of (CS/IT/IS) class 06 (database)
ESSENTIAL of (CS/IT/IS) class 06 (database)ESSENTIAL of (CS/IT/IS) class 06 (database)
ESSENTIAL of (CS/IT/IS) class 06 (database)
 
Model Call Girl in Tilak Nagar Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Tilak Nagar Delhi reach out to us at 🔝9953056974🔝Model Call Girl in Tilak Nagar Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Tilak Nagar Delhi reach out to us at 🔝9953056974🔝
 
9953330565 Low Rate Call Girls In Rohini Delhi NCR
9953330565 Low Rate Call Girls In Rohini  Delhi NCR9953330565 Low Rate Call Girls In Rohini  Delhi NCR
9953330565 Low Rate Call Girls In Rohini Delhi NCR
 
Framing an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdf
Framing an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdfFraming an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdf
Framing an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdf
 
ECONOMIC CONTEXT - LONG FORM TV DRAMA - PPT
ECONOMIC CONTEXT - LONG FORM TV DRAMA - PPTECONOMIC CONTEXT - LONG FORM TV DRAMA - PPT
ECONOMIC CONTEXT - LONG FORM TV DRAMA - PPT
 
CARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxCARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptx
 
Roles & Responsibilities in Pharmacovigilance
Roles & Responsibilities in PharmacovigilanceRoles & Responsibilities in Pharmacovigilance
Roles & Responsibilities in Pharmacovigilance
 
Model Call Girl in Bikash Puri Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Bikash Puri  Delhi reach out to us at 🔝9953056974🔝Model Call Girl in Bikash Puri  Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Bikash Puri Delhi reach out to us at 🔝9953056974🔝
 
Proudly South Africa powerpoint Thorisha.pptx
Proudly South Africa powerpoint Thorisha.pptxProudly South Africa powerpoint Thorisha.pptx
Proudly South Africa powerpoint Thorisha.pptx
 
Employee wellbeing at the workplace.pptx
Employee wellbeing at the workplace.pptxEmployee wellbeing at the workplace.pptx
Employee wellbeing at the workplace.pptx
 
“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...
“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...
“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...
 
Painted Grey Ware.pptx, PGW Culture of India
Painted Grey Ware.pptx, PGW Culture of IndiaPainted Grey Ware.pptx, PGW Culture of India
Painted Grey Ware.pptx, PGW Culture of India
 
Biting mechanism of poisonous snakes.pdf
Biting mechanism of poisonous snakes.pdfBiting mechanism of poisonous snakes.pdf
Biting mechanism of poisonous snakes.pdf
 
18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf
18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf
18-04-UA_REPORT_MEDIALITERAСY_INDEX-DM_23-1-final-eng.pdf
 
How to Configure Email Server in Odoo 17
How to Configure Email Server in Odoo 17How to Configure Email Server in Odoo 17
How to Configure Email Server in Odoo 17
 
Presiding Officer Training module 2024 lok sabha elections
Presiding Officer Training module 2024 lok sabha electionsPresiding Officer Training module 2024 lok sabha elections
Presiding Officer Training module 2024 lok sabha elections
 
Historical philosophical, theoretical, and legal foundations of special and i...
Historical philosophical, theoretical, and legal foundations of special and i...Historical philosophical, theoretical, and legal foundations of special and i...
Historical philosophical, theoretical, and legal foundations of special and i...
 
Computed Fields and api Depends in the Odoo 17
Computed Fields and api Depends in the Odoo 17Computed Fields and api Depends in the Odoo 17
Computed Fields and api Depends in the Odoo 17
 

CT Dose Issues.pptx on the factors to be considered on radiation protection

  • 2. Radiation Radiation is the emission and transmission of energy or particles of matter from atoms that travels through some material / through a vacuum1. 2 Types of Radiation 1. Particle Radiation (α & β radiation) 2. Electromagnetic Radiation (X and Gamma radiation)
  • 4. • Radiation Exposure: It describes the amount of radiation traveling through the air. Many radiation monitors measure exposure. The units for exposure are the roentgen (R)2 . • Absorbed Dose: which describes the amount of radiation energy deposited in the patient’s body as a result of exposure. The units for absorbed dose are the radiation absorbed dose (rad) and gray (Gy)2. • Effective Dose: are calculated by calculating the absorbed doses to individual organs weighted for their radiation sensitivity. • The radiation risk to the patient can be estimated from the effective dose. • The common unit to measure effective dose is the mSv 2.
  • 6. Stochastic Effects: • Stochastic effects are those that occur by chance and consist primarily of cancer and genetic effects3. • Stochastic effects often show up years after exposure3. • As the dose to an individual increases, the probability that cancer or a genetic effect will occur also increases3. • There is no threshold dose below which it is relatively certain that an adverse effect cannot occur for Stochastic Effects 1.
  • 7. Non Stochastic (Deterministic ) Effects: • These are characterized by a threshold dose above which they do occur (clear relationship between the exposure and the effect)1. • In addition; the magnitude of the effect is directly proportional to the size of the dose4. • Non stochastic effects typically result when very large dosages of radiation are received in a short amount of time4. • Examples of non stochastic effects include erythema (skin reddening), skin and tissue burns, cataract formation, sterility, radiation sickness and death4.
  • 8. CT’s Contribution X-Ray Procedure Radiation Dose CT: 49% Rest: 51% Total X-Ray Procedures CT:, 17,% Rest:, 83,%
  • 9. Estimated dosages of X-Ray Procedures Background radiation per annum ≈ 3mSv 1 CT Head ≈ 20 Skull X-ray Examinations 1 CT Abdomen/Pelvis (2 Phase Study) ≈ 32 Abdominal X-rays 1 CT Abdomen/Pelvis (Urogram) ≈ 60 Abdominal X-rays 1 CT Chest ≈ 70 Chest X-ray Examinations Radiograph Effective Dose (mSv) DEXA Scan 0.001 Bite Wing Dental 0.004 Skull (AP&Lat) 0.1 Chest (AP&Lat) 0.1 Abdomen (AP) 0.7 T-Spine (AP&Lat) 1 L-Spine (AP&Lat) 1.5 CT Procedure Effective Dose (mSv) Head 2 Neck 3 Chest 7 Abdomen 8 Pelvis 6 T-Spine 6 Chest for PE 15 Coronary Angio 16
  • 10. Factors to consider 4 Justification : Exam must be medically • indicated. 5 Optimization: Do the least number of scans at the lowest exposure . 6 Are non ionizing alternative better suited (MR and Ultrasound)5. 1 Patient Age. 2 Patient Gender. 3 Patient size and weight.
  • 11. CT Dose Descriptors 1. CT Dose Index (CTDIvol) ●This is the primary dose concept in MDCT measured in milli Gray (mGy). ● It is directly displayed on the MDCT Scanner. ●CTDIvol (weighted average of dose across a single slice) divided by a pitch factor. ●It can be used to directly compare radiation dose from different scan parameter settings. But it doses not provide an accurate dose to the patient and should not be used to determine individual risk5.
  • 12. CT Dose Descriptors 2. Dose Length Product (DLP) ●DLP is an indicator of the total dose for an entire CT examination. ● DLP is determined by multiplying the CTDIvol value by the scan length, (mGy-cm). ●DLP will change between patients due to differences in their height and the resulting scan length. ●It also includes overscan (extra Scanning from MDCT Scans). ●DLP alone is still not sufficient to accurately asses the radiation risk to a patient undergoing a MDCT scan5.
  • 13. CT Dose Descriptors 3. Effective Dose (ED) ●ED reflects the radiation risk to the individual patient, measured in milli Sievert (mSv). ●ED takes into account not only the amount of radiation, but also the sensitivity of the tissue/organ being radiated. ● ED = k x DLP . k is the weighting factor accounting for tissue sensitivity. ●Still mathematical models to compute organ doses are based on a standard adult (70kg) . Thus ED estimation can underestimate the risk to children/ thin patients and overestimate the risk to obese patients5.
  • 15. 1.Tube Current (mA) ●Tube Current in CT implies the amount of x-rays produced in the x-ray tube. ●mA and mAs has a linear relationship with respect to CT radiation dose. ●When all other technical factors are kept the same, and mAs are lowered , the radiation dosed is also lowered6.
  • 16. Factors Affecting Radiation Dose 2. Peak Tube Voltage (kVp) ●This is the peak potential difference between the anode and cathode in the x-ray tube. ●It cause acceleration of the electrons toward the anode to produce x-rays. ● The most common range of kVps used in CT are 80 -140. ●The output of an x-ray tube is approximately proportional to kVp2 . Therefore if kVp is doubled the radiation dose increase by a factor of 4 5.
  • 17. Factors Affecting Radiation Dose 3. Pitch ●Pitch is defined as the ratio of table feed per gantry rotation to total x-ray beam with. ● Radiation doses is inversely Proportional to the pitch. ● Therefore the greater the pitch the lower the radiation dose to the patient and vise-versa. ●Protocols demanding high spatial resolution (e.g. Cardiac ct scans) an pitch < 1 is used, translating in high radiation doses to the patient5.
  • 18. Factors Affecting Radiation Dose 4. Filters ●Consists of absorbent material inserted between the x-ray tube and the patient . ● Filters absorb low energy x-rays and “harden” the x-ray beam. ● Low energy x-rays contribute disproportionally to patient radiation dose (especially to skin dose). ●Bow tie filters (often used in CT) deliver a uniform intensity x-ray beam across the scanning object, resulting in a more uniform distribution of the dose5.
  • 19. Factors Affecting Radiation Dose 5. Patient Centering ●Patients normally have an elliptical cross- Section( thinner in the periphery and ticker in the centre. ● Bowtie filters are thicker in the periphery and thinner in the centre, resulting in more x-rays reaching the ticker central parts and less the thinner periphery. ● However for these filters to function properly the patient must be centred at the gantry iso-center. ●If the patient is positioned off centred less x-rays pass through the patient’s thicker central part ,resulting in more noise in the centre of the images. While more x-rays reach the peripheral portion of the body cross-section, increasing unwanted radiation dose6.
  • 20. Factors Affecting Radiation Dose 6. Repeats ●Multiple Scans results in an increase in radiation dose to the patient . ● During a single visit to a CT suite, the patient may be scanned multiple times. ● Patients can also undergo multiple ct scans within a short period of time. ●How many of these scans are necessary and appropriate Is the critical question that needs to be answered5.
  • 21. Factors Affecting Radiation Dose 7. Dose efficiency (Geometric efficiency) ●In MDCT it is required that an equal number of x-ray photons reach all active detectors. ●To ensure this a certain amount of over- beaming, beyond the active detectors are needed (called the penumbra). ●This proportion of wasted radiation and subsequent dose penalty is more profound with ct protocols using a thinner detector configuration than those using a thicker one . ●The narrower the beam and the less the active detector elements , the more radiation dose will increase5.
  • 22. Factors Affecting Radiation Dose 8. Angular (X/Y) Dynamic Dose Modulation ●It modulates the x-ray tube current during the scan according to the size of the scanned object to provide. ●This is to provide optimal penetration of the x-ray beam at a specific tube angle for a specific object. ●It provides the best dose saving for structure that are not uniform in the x/y planes. ●Provide dose saving of approximately 10% for head scans and 20-40% for chest scans6.
  • 23. Factors Affecting Radiation Dose 9. Longitudinal (Z-axis) Dose Modulation ●It modulates the mAs along the longitudinal (Z) axis. ●Information for this modulation are obtained from the surview (scout) scan. ● It reduce dose in organs that have lower attenuation(e.g: lungs) due to the lower dose needed to obtain diagnostic images. ●It also maintains an uniform image quality within the scan5.
  • 24. Factors Affecting Radiation Dose 10. Cardiac ECG Dose Modulation ● Retrospectively gated cardiac cta’s. - Higher radiation dose to patient. - More information that may be diagnostically relevant. - Give functional information (e.g. Ventricular ejection fractions) ● Prospectively gated cardiac cta’s. - Less radiation dose to patient (up to 83% less than retrospective studies). - Loss of information that may be diagnostically relevant. - No functional information (e.g. Ventricular ejection fractions)
  • 25. Factors Affecting Radiation Dose 11. Paediatric Protocols ●The goal is to obtain a balance between image quality and dose. ● These scan must be done with the minimum mA that will still provide a diagnostic relevant set of images. ● Low kVp settings in patients <45kg can result in 30-70% dose reduction. ●Doubling the pitch from 1 to 2 will result in a 50% decrease in the radiation dose. ●Paediatric examinations should also be performed with the shortest possible scan time (0.5 ms or less) to minimize patient movement and thus avoid repeating scan procedures7.
  • 26. Smart Card Protection for Patients ●Up until a decade ago, radiation protection was centred around protecting staff at medical facilities. ●Patient protection was seen as not so important, seen that their exposure to radiation was thought to be only a few times in a life time. ●This notion has certainly change, and the radiation dose patients received have become more prominent. ●What's needed is a way to track/record the radiation doses an individual received from medical facilities over his lifetime. ●To accomplish this the IAEA has launched a smart card
  • 27. REFERENCES 1. Health Physics Society . 2013. http://hps.org/publicinformation/ate/faqs/whatisradiation.html [2 June 2014]. 2. United States Nuclear Regulatory Commission. 2013. http://www.nrc.gov/about- nrc/radiation/health-effects/measuring-radiation.html [02 June 2014]. 3. United States Environmental Protection Agency . 2012. http://www.epa.gov/rpdweb00/understand/health_effects.html [02 June 2014]. 4. NDT Resource Center. 2013. https://www.nde- ed.org/EducationResources/CommunityCollege/RadiationSafety/biological/nonstochastic/nonst ochastic.htm [02 June 2014] 5. Mahesh, M. 2009. MDCT Physics: The Basics-Technology, Image Quality and Radiation Dose. Philadelphia: Lippincott Williams & Wilkins. 6. Kalra, M.K., Saini, S. & Rubin, G.D. 2008. MDCT from Protocols to Practice. Milan, Italy: Springer-Verslag Italia. 7. Siegal, M.J. 2007. Pediatric Body CT. 2nd ed. Philadelphia: Lippincott Williams & Wilkins. 8. International Atomic Energy Agency . http://www.iaea.org/Publications/Magazines/Bulletin/Bull502/50205813137.html [02 June 2014]