The document discusses the commissioning and quality assurance procedures needed for radiation treatment planning systems, including performing various dose calculation and distribution tests on photon and electron beams to validate the system is working accurately before clinical use. It also describes the ongoing quality control checks that should be done such as reproducibility tests, plan verification, and in vivo dosimetry to ensure continued safe and proper system performance. Commissioning a new system typically takes months while quality assurance should involve trained staff spending about 20 minutes checking each patient's plan.
Call Us ≽ 9953322196 ≼ Call Girls In Mukherjee Nagar(Delhi) |
RT10_EBT3c_GoodPractice_Planning.ppt
1. Radiation Protection in
Radiotherapy
Part 10
Good Practice including Radiation
Protection in EBT
Lecture 3 (cont.): Radiotherapy Treatment Planning
IAEA Training Material on Radiation Protection in Radiotherapy
2. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 2
C. Commissioning
Complex procedure depending very much on
equipment
Protocols exist and should be followed
Useful literature:
J van Dyk et al. 1993 Commissioning and QA of treatment
planning computers. Int. J. Radiat. Oncol. Biol. Phys. 26: 261-273
J van Dyk et al, 1999 Computerised radiation treatment planning
systems. In: Modern Technology of Radiation Oncology (Ed.: J
Van Dyk) Chapter 8. Medical Physics Publishing, Wisconsin,
ISBN 0-944838-38-3, pp. 231-286.
3. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 3
Acceptance testing and
commissioning
Acceptance testing: Check that the system conforms with
specifications.
Documentation of specifications either in the tender, in
guidelines or manufacturers’ notes – may test against
standard data (e.g. Miller et al. 1995, AAPM report 55)
Subset of commissioning procedure
Takes typically two weeks
Commissioning: Getting the system ready for clinical use
Takes typically several months for modern 3D system
4. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 4
Some equipment required
Scanning beam data acquisition system
Calibrated ionization chamber
Slab phantom including
inhomogeneities
Radiographic film
Anthropomorphic phantom
Ruler, spirit level
5. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 5
Commissioning
A. Non-dose related components
B. Photon dose calculations
C. Electron dose calculations
(D. Brachytherapy - covered in part 11)
E. Data transfer
F. Special procedures
6. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 6
A. Non-dose components
Image input
Geometry and scaling of
Digitizer,
Scans
Output
Text information
Anatomical structure information
CT numbers
Structures (outlining tools, non-axial
reconstruction, “capping”,…)
7. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 7
Electron and photon beams
Description (machine, modality, energy)
Geometry (Gantry, collimator, table,
arcs)
Field definition (Collimator, trays, MLC,
applicators, …)
Beam modifiers (Wedges, dynamic
wedges, compensators, bolus,…)
Normalization
8. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 9
B. Photon calculation tests
Point doses
TAR, TPR, PDD, PSF
Square, rectangular and irregular fields
Inverse square law
Attenuation factors (trays, wedges,…)
Output factors
Machine settings
9. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 10
Photon calculation tests (cont.)
Dose distribution
Homogenous
Profiles (open and wedged)
SSD/SAD
Contour correction
Blocks, MLC, asymmetric jaws
Multiple beams
Arcs
Off axis (open and wedged)
Collimator/couch rotation
PTW waterphantom
10. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 11
Photon calculation tests (cont.)
Dose distribution
Inhomogeneous
Slab geometry
Other geometries
Anthropomorphic phantom
In vivo dosimetry at least for the
first patients
Following the incident in Panama, the IAEA
recommends a largely extended in vivo dosimetry
program to be implemented
11. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 12
C. Electron calculation
Similar to photons, however, additional:
Bremsstrahlung tail
Small field sizes require special consideration
Inhomogeneity has more impact
It is possible to use reference data for
comparison (Shui et al. 1992 “Verification
data for electron beam dose algorithms” Med.
Phys. 19: 623-636)
12. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 13
E. Data transfer
Pixel values, CT numbers
Missing lines
Patient/scan information
Orientation
Distortion, magnification
All needs verification!!!
13. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 14
F. Special procedures
Junctions
Electron abutting
Stereotactic procedures
Small field procedures (e.g. for eye
treatment)
IMRT
TBI, TBSI
Intraoperative radiotherapy
14. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 15
Sources of uncertainty
Patient localization
Imaging (resolution, distortions,…)
Definition of anatomy (outlines,…)
Beam geometry
Dose calculation
Dose display and plan evaluation
Plan implementation
15. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 16
Typical accuracy required (examples)
Square field CAX:
1%
MLC penumbra: 3%
Wedge outer beam:
5%
Buildup-region: 30%
3D inhomogeneity
CAX: 5%
From AAPM TG53
16. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 17
Typical accuracy required (examples)
Square field CAX:
1%
MLC penumbra: 3%
Wedge outer beam:
5%
Buildup-region: 30%
3D inhomogeneity
CAX: 5%
Note:
Uncertainties have
two components:
Dose (given in %)
Location (given in
mm)
17. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 18
Time and staff requirements for
commissioning (J Van Dyk 1999)
Photon beam: 4-7 days
Electron beam: 3-5 days
Brachytherapy: 1 day per source type
Monitor unit calculation: 0.3 days per
beam
18. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 19
Some ‘tricky’ issues
Dose Volume Histograms - watch sampling,
grid, volume determination, normalization
(1% volume represents still > 10E7 cells!)
Biological parameters - Tumour Control
Probability (TCP) and Normal Tissue
Complication Probability (NTCP) depend on
the model used and the parameters which
are available.
19. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 20
Commissioning summary
Probably the most complex task for RT
physicists - takes considerable time and training
Partial commissioning needed for system
upgrades and modification
Documentation and hardcopy data must be
included
Training is essential and courses are available
Independent check highly recommended
21. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 22
Superficial beam
HVL
Percentage depth dose (may be look up table)
Normalization point (typically the surface)
Scatter (typically back scatter) factor
Applicator and/or cone factor
Timer accuracy
On/off effect
Other effects which may affect dose (e.g. electron
contamination)
22. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 23
Quality Assurance of a treatment
planning system
QA is typically a subset of commissioning
tests
Protocols:
As for commissioning and:
M Millar et al. 1997 ACPSEM position paper.
Australas. Phys. Eng. Sci. Med. 20 Supplement
B Fraas et al. 1998 AAPM Task Group 53: QA for
clinical RT planning. Med. Phys. 25: 1773-1829
23. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 24
Aspects of QA (compare also
part 12 of the course)
Training - qualified staff
Checks against a benchmark -
reproducibility
Treatment verification
QA administration
Communication
Documentation
Awareness of procedures required
24. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 25
Quality Assurance
25. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 26
Quality Assurance
Check prescription
Hand calculation of
treatment time
26. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 27
Frequency of tests for planning (and
suggested acceptance criteria)
Commissioning and significant upgrades
See above
Annual:
MU calculation (2%)
Reference plan set (2% or 2mm)
Scaling/geometry input/output devices (1mm)
Monthly
Check sum
Some reference test sets
27. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 28
Frequency of tests (cont.)
Weekly
Input/output devices
Each time system is turned on
Check sum (no change)
Each plan
CT transfer - orientation?
Monitor units - independent check
Verify input parameters (field size, energy, etc.)
28. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 29
Treatment planning QA summary
Training most essential
Staying alert is part of QA
Documentation and reporting necessary
Treatment verification in vivo can play
an important role
30. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 31
Staff and time requirements
(source J. Van Dyk et al. 1999)
Reproducibility tests/QC: 1 week per
year
In vivo dosimetry: about 1 hour per
patient - aim for about 10% of patients
Manual check of plans and monitor
units: 20 minutes per plan
31. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 32
QA in treatment planning
The planning system
QA of the system
Plan of a patient
QA of the plan
32. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 33
QC of treatment plans
Treatment plan:
Documentation of
treatment set-up,
machine parameters,
calculation details,
dose distribution,
patient information,
record and verify
data
Consists typically of:
Treatment sheet
Isodose plan
Record and Verify
entry
Reference films
(simulator, DRR)
33. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 34
QC of treatment plans
Check plan for each patient prior to
commencement of treatment
Plan must be
Complete from prescription to set-up
information and dose delivery advise
Understandable by colleagues
Document treatment for future use
34. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 35
Who should do it?
Treatment sheet checking should involve
senior staff
It is an advantage if different professions
can be involved in the process
Reports must go to clinicians and the
relevant QA committee
35. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 36
Example for physics treatment sheet
checking procedure
1. Check prescription (energy/dose/fractionation is everything signed ?)
2. Check prescription and calculation page for consistency: Isocentric (SAD) or fixed distance (SSD) set-up ? Are all
necessary factors used? Check both,dose/fraction and number of fractions.
3. Check normalisation value (Plan or data sheets).
4. Check outline, separation and prescription depth.
5. Turn to treatment plan: Does it look ok ? Outline ? Bolus ? Isocentre placement and normalisation point ? Any concerns
regarding the use of algorithms near surfaces or inhomogeneities? Would you expect problems in planes not shown ?
Prescription ?
6. Check and compare with treatment sheet calculation page: treatment unit and type, field names, weighting, wedges,
blocks, field size (FS), focus surface distance (FSD), Tissue Air Ratio (TAR) (if isocentric treatment) - is this consistent
with entries in treatment log page?
7. Electrons only: …
8. Photons only: …
9. Check shadow tray factor, wedge factor. Are any other attenuation factors required (e.g. couch, headrest, table tray...) ?
10. Check inverse square law factor (in electron treatments: is the virtual FSD appropriate?)
11. Calculate monitor units. Is time entry ok ?
12. Check if critical organ (e.g. spinal cord, lens, scrotum) dose or hot spot dose is required. If so, is it calculated correctly ?
13. Suggest in vivo dosimetry measurements if appropriate. Sign calculation sheet (if everything is ok).
14. Compare results on calculation page with entries in treatment log.
15. Check diagram and/or set up description: is there anything else worth to consider ?
16. Sign top of treatment sheet (specify what parts where checked if not all fields were checked).
17. Contact planning staff if required. Sign off physics log book.
36. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 37
Example for physics treatment sheet
checking procedure
1. Check prescription (energy/dose/fractionation is
everything signed ?)
2. Check prescription and calculation page for
consistency: Isocentric (SAD) or fixed distance (SSD)
set-up ? Are all necessary factors used? Check
both,dose/fraction and number of fractions.
3. Check normalisation value (Plan or data sheets).
4. Check outline, separation and prescription depth.
5. Turn to treatment plan: Does it look ok ? Outline ?
Bolus ? Isocentre placement and normalisation point ?
Any concerns regarding the use of algorithms near
surfaces or inhomogeneities? Would you expect
problems in planes not shown ? Prescription ?
37. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 38
Example for physics treatment sheet
checking procedure (cont.)
6. Check and compare with treatment sheet calculation page:
treatment unit and type, field names, weighting, wedges,
blocks, field size (FS), focus surface distance (FSD), Tissue
Air Ratio (TAR) (if isocentric treatment) - is this consistent with
entries in treatment log page?
7. Electrons only: …
8. Photons only: …
9. Check shadow tray factor, wedge factor. Are any other
attenuation factors required (e.g. couch, headrest, table
tray...) ?
10. Check inverse square law factor (in electron treatments: is the
virtual FSD appropriate?)
11. Calculate monitor units. Is time entry ok ?
12. Check if critical organ (e.g. spinal cord, lens, scrotum) dose or
hot spot dose is required. If so, is it calculated correctly ?
38. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 39
Example for physics treatment sheet
checking procedure (cont.)
13. Suggest in vivo dosimetry measurements if
appropriate. Sign calculation sheet (if everything is
ok).
14. Compare results on calculation page with entries in
treatment log.
15. Check diagram and/or set up description: is there
anything else worth to consider ?
16. Sign top of treatment sheet (specify what parts
where checked if not all fields were checked).
17. Contact planning staff if required. Sign off
physics log book.
39. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 40
Treatment plan QA summary
Essential part of departmental QA
Part of patient records
Multidisciplinary approach
41. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 42
Did we achieve the objectives?
Understand the general principles of
radiotherapy treatment planning
Appreciate different dose calculation
algorithms
Be able to apply the concepts of optimization
of medical exposure throughout the treatment
planning process
Appreciate the need for quality assurance in
radiotherapy treatment planning
42. Radiation Protection in Radiotherapy Part 10, lecture 3 (cont.): Radiotherapy treatment planning 43
Overall Summary
Treatment planning is the most important step
towards radiotherapy for individual patients -
as such it is essential for patient protection as
outlined in BSS
Treatment planning is growing more complex
and time consuming
Understanding of the process is essential
QA of all aspects is essential